Philips Plus Computed Tomography System
System, X-Ray, Tomography, Computed. 510(k) K033326, from Philips Medical Systems, Inc. (Cleveland, OH). Product code JAK, class 2.
As of Nov 17, 2026, K033326 is an FDA 510(k) held by Philips Medical Systems, Inc., substantially equivalent on Oct 29, 2003 after 13 days, with 25 recalls tied to it in FDA's recall database.
- Substantially equivalent: Oct 29, 2003
- Review time: 13 days
- Recalls: 25+
- Cited as a predicate by: 9 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- MX8000 Idt Ct SystemK012009 · Marconi Medical Systems, Inc. · Sep 13, 2001
- MX8000 V5.0 Ct SystemK010817 · Philips Medical Systems (Cleveland), Inc. · Jun 15, 2001
Cited as a Predicate
9 later 510(k) summaries name this device as a predicate- uCT Computed Tomography X-ray SystemK172135 · Shanghai United Imaging Healthcare Co., Ltd. · Mar 2, 2018
- Philips Ingenuity CTK160743 · Philips Medical Systems (Cleveland), Inc. · Aug 8, 2016
- Philips Multislice CT System with Low Dose CT Lung CancerK153444 · Philips Medical Systems (Cleveland), Inc. · Apr 8, 2016
- NeuViz 128 Multi-slice CT Scanner SystemK151383 · Neusoft Medical Systems Co., Ltd. · Nov 4, 2015
- Modification To Neuviz Dual Multi-Slice Ct Scanner SystemK071308 · Neusoft Medical Systems Co., Ltd. · Aug 16, 2007
- Brilliance VolumeK060937 · Philips Medical Systems (Cleveland), Inc. · Jun 5, 2006
- Precedence Spect/Ct Imaging SystemK051351 · Adac Laboratories · Jun 8, 2005
- Gemini GxlK051170 · Philips Medical Systems (Cleveland), Inc. · May 23, 2005
- Brilliance Ct, Private Practice Cv Configuration, Ct ScannerK042293 · Philips Medical Systems (Cleveland), Inc. · Sep 8, 2004
Recalls
Recalls FDA ties to K033326, newest first; the 25 newest; each opens FDA's recall record| Aug 7, 2026 | Z-3043-2026Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment. | Open, Classified |
| Aug 13, 2025 | Z-2603-2025The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. | Open, Classified |
| Aug 13, 2025 | Z-2602-2025The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. | Open, Classified |
| Aug 13, 2025 | Z-2604-2025The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. | Open, Classified |
| Aug 13, 2025 | Z-2595-2025The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. | Open, Classified |
| May 1, 2021 | Z-1982-2021Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall… | Open, Classified |
| Oct 18, 2018 | Z-0544-2019There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall | Terminated |
| Oct 18, 2018 | Z-0543-2019There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall | Terminated |
| Jul 3, 2018 | Z-1143-2019Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the… | Terminated |
| Mar 30, 2018 | Z-2174-2018The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude… | Terminated |
| Feb 21, 2018 | Z-1396-2018Surview scan with tube current lower than 30 mA is unable to be initialized. | Terminated |
| Feb 8, 2018 | Z-1407-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities. | Terminated |
| Feb 8, 2018 | Z-1404-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities. | Terminated |
| Feb 8, 2018 | Z-1405-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities. | Terminated |
| Feb 8, 2018 | Z-1406-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities. | Terminated |
| Feb 8, 2018 | Z-1413-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities. | Terminated |
| Nov 10, 2017 | Z-0389-2018Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the… | Terminated |
| Nov 12, 2016 | Z-0694-2017Software error due to the filament on timer | Terminated |
| Nov 12, 2016 | Z-0693-2017Software error due to the filament on timer | Terminated |
| Aug 24, 2016 | Z-0009-2017The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may… | Terminated |
| Aug 24, 2016 | Z-0010-2017The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may… | Terminated |
| Aug 24, 2016 | Z-0012-2017The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may… | Terminated |
| Aug 24, 2016 | Z-0011-2017The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may… | Terminated |
| Jul 20, 2016 | Z-2658-2016After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that… | Terminated |
| Jul 20, 2016 | Z-2652-2016After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that… | Terminated |
Other JAK Clearances
The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Oct 16, 2003
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files