Brilliance Volume

System, X-Ray, Tomography, Computed. 510(k) K060937, from Philips Medical Systems (Cleveland), Inc. (Cleveland, OH). Product code JAK, class 2.

As of Nov 17, 2026, K060937 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on Jun 5, 2006 after 61 days, with 25 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 5, 2006
  • Review time: 61 days
  • Recalls: 25+
  • Cited as a predicate by: 10 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

10 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K060937, newest first; the 25 newest; each opens FDA's recall record
Aug 13, 2025Z-2599-2025The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.Open, Classified
May 1, 2021Z-1990-2021Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall…Open, Classified
Apr 13, 2021Z-1774-2021Cooling Unit (CLU) expelled out of a CT gantry of an iCT system may result in injuryTerminated
Apr 6, 2020Z-1921-2020The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.Terminated
Apr 6, 2020Z-1922-2020The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.Terminated
Feb 8, 2018Z-1409-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.Terminated
Feb 8, 2018Z-1408-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.Terminated
Nov 8, 2017Z-0331-2018There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard…Terminated
Nov 8, 2017Z-0332-2018There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard…Terminated
Jul 11, 2017Z-0403-2018Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image…Terminated
Jul 11, 2017Z-0402-2018Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image…Terminated
Dec 23, 2016Z-1058-2017During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.Terminated
Nov 25, 2016Z-1004-2017Failure to correctly document the installation of four M12 Bolts into the system rotor.Terminated
Oct 13, 2016Z-0593-2017The product label does not include the correct current rating.Terminated
Oct 13, 2016Z-0594-2017The product label does not include the correct current rating.Terminated
Jun 21, 2016Z-2112-2016Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack.Terminated
Apr 29, 2016Z-2382-2016Software issues found in v4.1 .3/4.1.5 in the Philips Brilliance iCT/ iCT SP products that could affect the performance of the equipment.Terminated
Apr 29, 2016Z-2381-2016Software issues in versions v4.1 .3/4.1.5 in the Philips Brilliance iCT) iCT SP products that, could affect the performance of the equipment.Terminated
Mar 4, 2016Z-1506-2016The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector…Terminated
Feb 5, 2016Z-1352-2016The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.Terminated
Feb 5, 2016Z-1353-2016The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.Terminated
Oct 29, 2015Z-0546-2016Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the…Terminated
Oct 29, 2015Z-0545-2016Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the…Terminated
Sep 6, 2015Z-1714-2016Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of…Terminated
Sep 6, 2015Z-1713-2016Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of…Terminated

Other JAK Clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional
Received
Apr 5, 2006
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files