Brilliance Volume
System, X-Ray, Tomography, Computed. 510(k) K060937, from Philips Medical Systems (Cleveland), Inc. (Cleveland, OH). Product code JAK, class 2.
As of Nov 17, 2026, K060937 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on Jun 5, 2006 after 61 days, with 25 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jun 5, 2006
- Review time: 61 days
- Recalls: 25+
- Cited as a predicate by: 10 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Philips Plus Computed Tomography SystemK033326 · Philips Medical Systems, Inc. · Oct 29, 2003
Cited as a Predicate
10 later 510(k) summaries name this device as a predicate- The Multimodality Advanced Vessel AnalysisK203216 · Philips Medical Systems Nederland B.V. · Nov 25, 2020
- Philips iCT CT SystemK162838 · Philips Medical Systems (Cleveland), Inc. · Apr 7, 2017
- I4 (Integrated Intelligent Imaging Informatics) systemK160315 · Philips Medical Systems Nederland B.V. · Feb 19, 2016
- Philips Multislice CT System with Low Dose CT Lung CancerK153444 · Philips Medical Systems (Cleveland), Inc. · Apr 8, 2016
- Philips Spectral CT ApplicationsK150665 · Philips Medical Systems Nethrlands BV · Aug 7, 2015
- Philips Ct Dynamic Myocardial Perfusion (Dmp) ApplicationK133603 · Philips Medical Systems · Sep 25, 2014
- Brilliance IctK131773 · Philips Healthcare (Cleveland) · Sep 24, 2013
- Imr Software ApplicationK123576 · Philips Healthcare (Cleveland) · Jun 7, 2013
- Iterative Reconstruction Technique Software ApplicationK113483 · Philips Healthcare (Cleveland) · Sep 26, 2012
- Brilliance Dual Energy OptionK090462 · Philips Medical Systems (Cleveland), Inc. · Jun 23, 2009
Recalls
Recalls FDA ties to K060937, newest first; the 25 newest; each opens FDA's recall record| Aug 13, 2025 | Z-2599-2025The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. | Open, Classified |
| May 1, 2021 | Z-1990-2021Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall… | Open, Classified |
| Apr 13, 2021 | Z-1774-2021Cooling Unit (CLU) expelled out of a CT gantry of an iCT system may result in injury | Terminated |
| Apr 6, 2020 | Z-1921-2020The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems. | Terminated |
| Apr 6, 2020 | Z-1922-2020The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems. | Terminated |
| Feb 8, 2018 | Z-1409-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities. | Terminated |
| Feb 8, 2018 | Z-1408-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities. | Terminated |
| Nov 8, 2017 | Z-0331-2018There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard… | Terminated |
| Nov 8, 2017 | Z-0332-2018There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard… | Terminated |
| Jul 11, 2017 | Z-0403-2018Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image… | Terminated |
| Jul 11, 2017 | Z-0402-2018Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image… | Terminated |
| Dec 23, 2016 | Z-1058-2017During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued. | Terminated |
| Nov 25, 2016 | Z-1004-2017Failure to correctly document the installation of four M12 Bolts into the system rotor. | Terminated |
| Oct 13, 2016 | Z-0593-2017The product label does not include the correct current rating. | Terminated |
| Oct 13, 2016 | Z-0594-2017The product label does not include the correct current rating. | Terminated |
| Jun 21, 2016 | Z-2112-2016Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack. | Terminated |
| Apr 29, 2016 | Z-2382-2016Software issues found in v4.1 .3/4.1.5 in the Philips Brilliance iCT/ iCT SP products that could affect the performance of the equipment. | Terminated |
| Apr 29, 2016 | Z-2381-2016Software issues in versions v4.1 .3/4.1.5 in the Philips Brilliance iCT) iCT SP products that, could affect the performance of the equipment. | Terminated |
| Mar 4, 2016 | Z-1506-2016The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector… | Terminated |
| Feb 5, 2016 | Z-1352-2016The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices. | Terminated |
| Feb 5, 2016 | Z-1353-2016The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices. | Terminated |
| Oct 29, 2015 | Z-0546-2016Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the… | Terminated |
| Oct 29, 2015 | Z-0545-2016Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the… | Terminated |
| Sep 6, 2015 | Z-1714-2016Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of… | Terminated |
| Sep 6, 2015 | Z-1713-2016Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of… | Terminated |
Other JAK Clearances
The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 5, 2006
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files