Gemini Gxl
System, Tomography, Computed, Emission. 510(k) K051170, from Philips Medical Systems (Cleveland), Inc. (Cleveland, OH). Product code KPS, class 2.
As of Nov 18, 2026, K051170 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on May 23, 2005 after 17 days, with 12 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 23, 2005
- Review time: 17 days
- Recalls: 12
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Gemini GxlK042935 · Philips Medical Systems (Cleveland), Inc. · Nov 9, 2004
- Philips Plus Computed Tomography SystemK033326 · Philips Medical Systems, Inc. · Oct 29, 2003
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Gemini RaptorK052640 · Philips Medical Systems (Cleveland), Inc. · Oct 7, 2005
Recalls
Recalls FDA ties to K051170, newest first; each opens FDA's recall record| May 1, 2021 | Z-0280-2022Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or… | Open, Classified |
| May 1, 2021 | Z-0278-2022Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or… | Open, Classified |
| Jul 1, 2015 | Z-2193-2015Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue. | Terminated |
| Mar 6, 2015 | Z-1781-2015The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or… | Terminated |
| Aug 8, 2012 | Z-2239-2012Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed… | Terminated |
| Jan 26, 2012 | Z-1368-2012TumorLOC will save the incorrect contour when the user saves the contour while in the edit mode, and the software does not force the user to exit the edit mode before saving. In TumorLOC, when saving while a contour is… | Terminated |
| Feb 19, 2009 | Z-1579-2009A software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software… | Terminated |
| Jan 26, 2009 | Z-0260-2010Upper patient pallet component of the recalling firm's Gemini PET/CT system may be thinner than normally required, which may reduce the product's load bearing capacity, resulting in cracking under less than the designed… | Terminated |
| Nov 24, 2008 | Z-0286-2010A leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images. | Terminated |
| May 1, 2008 | Z-0197-2016The reason for this inspection is to determine if the Spindle Block Servo Controllers on the CT subsystem of potentially affected GEMINI PET/CT systems are susceptible to premature failure. | Terminated |
| Sep 21, 2007 | Z-0462-2008Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment. | Terminated |
| Sep 21, 2007 | Z-0461-2008Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment. | Terminated |
Other KPS Clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7 (11364753);Symbia Pro.specta VA40…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI Panvivo (uMI Panvivo ES)K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series Digital SPECT/CT System…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- May 6, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files