System, Tomography, Computed, Emission. 510(k) K052640, from Philips Medical Systems (Cleveland), Inc. (Cleveland, OH). Product code KPS, class 2.
As of Nov 19, 2026, K052640 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on Oct 7, 2005 after 11 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 11 days
- Recalls: 25+
- Cited as a predicate by: 6 devices
The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data.
- Gemini GxlK051170 · Philips Medical Systems (Cleveland), Inc. · May 23, 2005
- Ingenuity TFK172406 · Philips Medical Systems (Cleveland), Inc. · Oct 6, 2017
- Celesteion, PCA-9000A/2K140651 · Toshibamedical Systems Corporation · Jun 11, 2014
- Ingenuity Digital Pet/CtK123599 · Philips Healthcare (Cleveland) · Jul 12, 2013
- Ingenuity TF Pet/MrK103483 · Philips Healthcare (Cleveland) · Nov 23, 2011
- Nexstar Liftoff Pet Application Software SuiteK081426 · Philips Medical Systems (Cleveland), Inc. · Jun 4, 2008
- Gemini CondorK081135 · Philips Medical Systems (Cleveland), Inc. · Apr 29, 2008
| May 1, 2021 | Z-0279-2022, Z-0281-2022Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or… | Open |
| May 1, 2021 | Z-1987-2021Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall… | Open |
| Mar 28, 2018 | Z-2276-2018A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users. | Terminated |
| Feb 15, 2018 | Z-1392-2018Following a period of inactivity, the mass storage device may cause the acquisition console to become unresponsive. This may prevent an acquisition from proceeding. | Terminated |
| Nov 1, 2017 | Z-0522-2018One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware. | Terminated |
| Jul 1, 2015 | Z-2195-2015, Z-2194-2015, Z-2198-2015, Z-2197-2015, Z-2196-2015Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue. | Terminated |
| Mar 6, 2015 | Z-1785-2015, Z-1782-2015, Z-1786-2015, Z-1783-2015, Z-1787-2015, Z-1788-2015The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or… | Terminated |
| Jun 18, 2014 | Z-2037-2014When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the… | Terminated |
| Dec 4, 2013 | Z-0932-2014The PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study. | Terminated |
| Aug 16, 2013 | Z-0067-2014Issues with the Ingenuity TF PET/CT Software Version 4.0 could pose a risk for patients in the event that a patient rescan were to require reinjection of PET radiopharmaceutical. | Terminated |
| Jan 21, 2013 | Z-0897-2013The system may not default to the same gating trigger with which the data were acquired. This could cause a risk for patients. | Terminated |
| Dec 20, 2012 | Z-1029-2013Issues have been detected in the Ingenuity TF PET/CT currently running software version 4.0.60605.0 that if they were to re-occur, could cause a risk for patients, users or service personnel, including inaccurate… | Terminated |
| Oct 24, 2012 | Z-0882-2013Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A… | Terminated |
| Aug 8, 2012 | Z-2239-2012Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed… | Terminated |
| Jan 26, 2012 | Z-1369-2012TumorLOC will save the incorrect contour when the user saves the contour while in the edit mode, and the software does not force the user to exit the edit mode before saving. In TumorLOC, when saving while a contour is… | Terminated |
- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7 (11364753);Symbia Pro.specta VA40…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI Panvivo (uMI Panvivo ES)K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series Digital SPECT/CT System…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Sep 26, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files