MX8000 Idt Ct System

System, X-Ray, Tomography, Computed. 510(k) K012009, from Marconi Medical Systems, Inc. (Cleveland, OH). Product code JAK, class 2.

As of Nov 18, 2026, K012009 is an FDA 510(k) held by Marconi Medical Systems, Inc., substantially equivalent on Sep 13, 2001 after 78 days, with 25 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 13, 2001
  • Review time: 78 days
  • Recalls: 25+
  • Cited as a predicate by: 12 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

12 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K012009, newest first; the 25 newest; each opens FDA's recall record
Aug 13, 2025Z-2596-2025The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.Open, Classified
May 1, 2021Z-1981-2021Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall…Open, Classified
Oct 18, 2018Z-0547-2019There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fallTerminated
Oct 18, 2018Z-0545-2019There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fallTerminated
Oct 18, 2018Z-0548-2019There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fallTerminated
Oct 18, 2018Z-0546-2019There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fallTerminated
Jul 3, 2018Z-1142-2019Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the…Terminated
Mar 30, 2018Z-2176-2018The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude…Terminated
Mar 30, 2018Z-2175-2018The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude…Terminated
Feb 8, 2018Z-1415-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.Terminated
Feb 8, 2018Z-1412-2018Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.Terminated
Jul 20, 2016Z-2656-2016After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that…Terminated
Jul 20, 2016Z-2655-2016After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that…Terminated
Mar 24, 2016Z-2369-2016Software defect causing intermittently slow response of Host.Terminated
Mar 24, 2016Z-2370-2016Software defect causing intermittently slow response of Host.Terminated
Mar 4, 2016Z-1506-2016The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector…Terminated
Sep 14, 2015Z-0173-2016The firm discovered Ring/Dot artifact due to X-ray measurement error.Terminated
Mar 16, 2015Z-1652-2015The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.Terminated
Mar 16, 2015Z-1653-2015The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.Terminated
Mar 16, 2015Z-1654-2015The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.Terminated
Mar 16, 2015Z-1651-2015The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.Terminated
Mar 16, 2015Z-1655-2015The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.Terminated
Mar 6, 2015Z-1766-2015The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or…Terminated
Mar 6, 2015Z-1765-2015The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or…Terminated
Mar 6, 2015Z-1793-2015The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or…Terminated

Other JAK Clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional
Received
Jun 27, 2001
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.