Gemini Pet/Ct Imaging System
System, Tomography, Computed, Emission. 510(k) K041955, from Philips Medical Systems (Cleveland), Inc. (Cleveland, OH). Product code KPS, class 2.
As of Nov 18, 2026, K041955 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on Jul 29, 2004 after 8 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 29, 2004
- Review time: 8 days
- Recalls: 2
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Gemini 16K032036 · Philips Medical Systems (Cleveland), Inc. · Jul 11, 2003
- MX8000 Idt Ct SystemK012009 · Marconi Medical Systems, Inc. · Sep 13, 2001
- Allegro Imaging SystemK033782 · Philips Medical Systems (Cleveland), Inc. · Dec 19, 2003
- Modification To Ge Discovery Ls SystemK040172 · GE Medical Systems · Apr 7, 2004
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- Ep Navigator SoftwareK062650 · Philips Medical Systems · Sep 21, 2006
- Gemini GxlK042935 · Philips Medical Systems (Cleveland), Inc. · Nov 9, 2004
Recalls
Recalls FDA ties to K041955, newest first; each opens FDA's recall record| Mar 6, 2015 | Z-1776-2015The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or… | Terminated |
| Nov 6, 2006 | Z-0013-2007A potential anomaly was identified on the CT subsytem, which could result in the radiation dose being different than the calculated dose reported to the operator. | Terminated |
Other KPS Clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7 (11364753);Symbia Pro.specta VA40…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI Panvivo (uMI Panvivo ES)K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series Digital SPECT/CT System…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Jul 21, 2004
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files