Philips Ingenuity CT

System, X-Ray, Tomography, Computed. 510(k) K160743, from Philips Medical Systems (Cleveland), Inc. (Cleveland, OH). Product code JAK, class 2.

As of Nov 19, 2026, K160743 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on Aug 8, 2016 after 144 days, with 24 recalls tied to it in FDA's recall database.

  • Review time: 144 days
  • Recalls: 24
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Ingenuity CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K160743, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 7, 2026Z-3042-2026Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.Open
Jun 11, 2026Z-2652-2026The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being…Open
Aug 13, 2025Z-2600-2025The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.Open
Jun 8, 2021Z-2315-2021, Z-2316-2021When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the…Terminated
May 1, 2021Z-1984-2021Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall…Open
Jun 24, 2019Z-2188-2019, Z-2186-2019, Z-2187-2019In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been appliedTerminated
Mar 28, 2018Z-2274-2018, Z-2277-2018, Z-2275-2018A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.Terminated
Nov 10, 2017Z-0390-2018, Z-0391-2018Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the…Terminated
Oct 6, 2017Z-0521-2018, Z-0518-2018, Z-0520-2018, Z-0519-2018Numerous issues related to software Brilliance iCT 4.1.6 software version.Terminated
Nov 12, 2016Z-0695-2017, Z-0697-2017, Z-0696-2017Software error due to the filament on timerTerminated
Oct 13, 2016Z-0597-2017, Z-0595-2017, Z-0596-2017The product label does not include the correct current rating.Terminated

Other JAK clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional
Received
Mar 17, 2016
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files