Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84…

Recall Z-1656-2025, part of FDA recall event 96567. See Abbott Laboratories's company page.

Abbott Laboratories started this Class II device recall on Mar 18, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Mar 18, 2025
  • Status: Ongoing
  • Products in this event: 5

Why It Was Recalled

Class II

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Record

Product
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby
Quantity
45
Lots and codes
UDI-DI 00380740017170 SNs: 35939BG 35979BG 36019BG 36103BG 36115BG 36143BG 36171BG 36177BG 36361BG 36788BG 36866BG 54095BG 54431BG 54686BG 54787BG 54900BG 55634BG 55644BG 55656BG 55657BG 55659BG 55662BG 55665BG 55672BG 55673BG 55747BG 55859BG 55879BG 70021BG 70161BG 70204BG 70463BG 70713BG 70931BG 71185BG 71387BG 71388BG 71397BG 71398BG 71476BG 71503BG 71626BG 71637BG 71638BG 71710BG
Distribution
Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.
Recalling firm
Abbott Laboratories, Abbott Park, IL, United States
How the firm notified
Letter
FDA classified
Apr 22, 2025