Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01…

Recall Z-1652-2025, part of FDA recall event 96567. See Abbott Laboratories's company page.

Abbott Laboratories started this Class II device recall on Mar 18, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Mar 18, 2025
  • Status: Ongoing
  • Products in this event: 5

Why It Was Recalled

Class II

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Record

Product
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Quantity
2947
Lots and codes
UDI-DI 00380740017170 SNs: 30039BG 34021BG 34079BG 34108BG 34123BG 34138BG 34140BG 34149BG 34152BG 34161BG 34177BG 34186BG 34232BG 34242BG 34272BG 34288BG 34302BG 34323BG 34339BG 34349BG 34353BG 34354BG 34364BG 34372BG 34391BG 34395BG 34414BG 34433BG 34439BG 34448BG 34451BG 34458BG 34466BG 34468BG 34473BG 34478BG 34488BG 34491BG 34531BG 34533BG 34539BG 34551BG 34555BG 34559BG 34565BG 34566BG 34598BG 34604BG 34605BG 34620BG 34633BG 34634BG 34646BG 34667BG 34669BG 34695BG 34726BG 34734BG 34736BG 34742BG 34744BG 34758BG 34768BG 34788BG 34789BG 34791BG 34793BG 34805BG 34816BG 34837BG 34839BG…
Distribution
Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.
Recalling firm
Abbott Laboratories, Abbott Park, IL, United States
How the firm notified
Letter
FDA classified
Apr 22, 2025

Other Products in This Recall

Recall event 96567

Abbott Laboratories's Other Recalls

Newest first, one product per recall event; search them all