Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number:…
Recall Z-0489-2026, part of FDA recall event 97788. See Abbott Laboratories's company page.
Abbott Laboratories started this Class II device recall on Sep 18, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 18, 2025
- Status: Ongoing
Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
- Product
- Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS, 5.7 and 5.10.1 in the US Product Description: Alinity hq Analyzer Component: No
- Quantity
- 833
- Lots and codes
- SN UDI HQ00480 (01)380740138851(21)HQ00480 HQ01231 (01)380740138851(21)HQ01231 HQ01232 (01)380740138851(21)HQ01232 HQ01233 (01)380740138851(21)HQ01233 HQ01236 (01)380740138851(21)HQ01236 HQ01237 (01)380740138851(21)HQ01237 HQ01238 (01)380740138851(21)HQ01238 HQ01239 (01)380740138851(21)HQ01239 HQ01240 (01)380740138851(21)HQ01240 HQ01241 (01)380740138851(21)HQ01241 HQ01244 (01)380740138851(21)HQ01244 HQ01246 (01)380740138851(21)HQ01246 HQ01247 (01)380740138851(21)HQ01247 HQ01248 (01)380740138851(21)HQ01248 HQ01249 (01)380740138851(21)HQ01249 HQ01250 (01)380740138851(21)HQ01250 HQ01252…
- Distribution
- Worldwide distribution: US (nationwide) to states of: AL, CA, CO, FL, IN, MN, NC, NY, OK and OUS (foreign) to countries of: ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAMAICA, JAPAN, JORDAN, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MALDIVES, MEXICO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SERBIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE,UNITED KINGDOM, URUGUAY, and VIETNAM.
- Recalling firm
- Abbott Laboratories, Santa Clara, CA, United States
- How the firm notified
- —
- FDA classified
- Nov 12, 2025
Newest first, one product per recall event; search them all
- Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20Z-0961-2026 · Class II · Nov 13, 2025 · Ongoing
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-03 Software Version: All software…Z-1653-2025 · Class II · Mar 18, 2025 · Ongoing
- Alinity hq Analyzer, REF: 09P68-01Z-2681-2024 · Class II · Jun 14, 2024 · Ongoing
- ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.Z-1476-2024 · Class II · Feb 21, 2024 · Ongoing
- Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module is a fully automated chemistry analyzer…Z-0681-2024 · Class II · Oct 19, 2023 · Ongoing
- Abbott CELL-DYN Ruby, Model CD-Ruby, List number 08H67-01, and Abbott CELL-DYN Sapphire, Model CD Sapphire, list number…Z-0604-2024 · Class II · Oct 19, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports