Arca
Skip to content
Tools
Tools

Alinity ci-series System Control Module (SCM)

Recall Z-0961-2026, part of FDA recall event 98031. See Abbott Laboratories's company page.

Abbott Laboratories started this Class II device recall on Nov 13, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Nov 13, 2025
  • Status: Ongoing

Why It Was Recalled

Class II

Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance issues found in the Alinity ci-series System software versions 3.6.1. and lower that could lead to erroneous results for multiple analytes. The issue was identified by Abbott during the internal testing of complaint investigations.

Record

Product
Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20
Quantity
11653
Lots and codes
UDI-DI: 00380740137410. GTIN: 06952999490048. Software versions 3.6.1 and lower. Serial Numbers: SCM01010 SCM01012 SCM01016 SCM01035 SCM01043 SCM01051 SCM01063 SCM01069 SCM01075 SCM01076 SCM01078 SCM01082 SCM01085 SCM01086 SCM01087 SCM01088 SCM01092 SCM01093 SCM01094 SCM01095 SCM01098 SCM01099 SCM01101 SCM01102 SCM01106 SCM01107 SCM01110 SCM01115 SCM01116 SCM01120 SCM01121 SCM01122 SCM01123 SCM01124 SCM01126 SCM01127 SCM01128 SCM01129 SCM01132 SCM01134 SCM01135 SCM01136 SCM01137 SCM01138 SCM01142 SCM01143 SCM01149 SCM01150 SCM01151 SCM01154 SCM01163 SCM01166 SCM01170 SCM01171 SCM01172…
Distribution
Worldwide distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NH, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR; and OUS (foreign) to countries of: Albania, Algeria, Andorra, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cambodia, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, Gaza & Jericho, Georgia, Germany, Greece, Greenland, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malawi, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Montenegro, Morocco, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Northern Ireland, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia (CIS), Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad & Tobago, Tunisia, Turkey, UAE, Uganda, Ukraine, United Kingdom, Uruguay, Vietnam and Zambia. OUS: ALBANIA, ALGERIA, ANDORRA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BRUNEI, BULGARIA, CAMBODIA, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FAROE ISLANDS, FINLAND, FRANCE, GAZA & JERICHO, GEORGIA, GERMANY, GREECE, GREENLAND, GUATEMALA, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOSOVO, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, MACEDONIA, MALAWI, MALAYSIA, MALDIVES, MALTA, MAURITANIA, MAURITIUS, MEXICO, MOLDOVA, MONTENEGRO, MOROCCO, NAMIBIA, NEPAL, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, NORTHERN IRELAND, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA (CIS), RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TUNISIA, TURKEY, UAE, UGANDA, UKRAINE, UNITED KINGDOM, URUGUAY, VIETNAM, ZAMBIA
Recalling firm
Abbott Laboratories, Irving, TX, United States
How the firm notified
—
FDA classified
Dec 18, 2025

Abbott Laboratories's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc