Alinity hq Analyzer, REF: 09P68-01
Recall Z-2681-2024, part of FDA recall event 94958. See Abbott Laboratories's company page.
Abbott Laboratories started this Class II device recall on Jun 14, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jun 14, 2024
- Status: Ongoing
1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts
- Product
- Alinity hq Analyzer, REF: 09P68-01
- Quantity
- 816
- Lots and codes
- UDI-DI: 380740138851. Software Versions: 5.7 (US) and 5.6 and below (OUS), Serial Numbers: HQ00105, HQ00108, HQ00117, HQ00118, HQ00119, HQ00120, HQ00126, HQ00127, HQ00128, HQ00132, HQ00136, HQ00139, HQ00140, HQ00141, HQ00142, HQ00144, HQ00145, HQ00147, HQ00148, HQ00149, HQ00152, HQ00153, HQ00154, HQ00156, HQ00157, HQ00161, HQ00162, HQ00166, HQ00169, HQ00171, HQ00172, HQ00173, HQ00174, HQ00175, HQ00183, HQ00184, HQ00185, HQ00186, HQ00187, HQ00188, HQ00266, HQ00268, HQ00269, HQ00270, HQ00271, HQ00273, HQ00274, HQ00276, HQ00277, HQ00278, HQ00279, HQ00281, HQ00282, HQ00283, HQ00286, HQ00287,…
- Distribution
- Worldwide - US Nationwide distribution in the states of NY, OK, NC and the countries of ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, Botswana, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAMAICA, JAPAN, JORDAN, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MALDIVES, MEXICO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SERBIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, URUGUAY, VIETNAM.
- Recalling firm
- Abbott Laboratories, Santa Clara, CA, United States
- How the firm notified
- Letter
- FDA classified
- Aug 16, 2024
- Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20Z-0961-2026 · Class II · Nov 13, 2025 · Ongoing
- Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version:…Z-0489-2026 · Class II · Sep 18, 2025 · Ongoing
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-03 Software Version: All software…Z-1653-2025 · Class II · Mar 18, 2025 · Ongoing
- ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.Z-1476-2024 · Class II · Feb 21, 2024 · Ongoing
- Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module is a fully automated chemistry analyzer…Z-0681-2024 · Class II · Oct 19, 2023 · Ongoing
- Abbott CELL-DYN Ruby, Model CD-Ruby, List number 08H67-01, and Abbott CELL-DYN Sapphire, Model CD Sapphire, list number…Z-0604-2024 · Class II · Oct 19, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports