Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-03…
Recall Z-1653-2025, part of FDA recall event 96567. See Abbott Laboratories's company page.
Abbott Laboratories started this Class II device recall on Mar 18, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Mar 18, 2025
- Status: Ongoing
- Products in this event: 5
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
- Product
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-03 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
- Quantity
- 146
- Lots and codes
- UDI-DI 00380740017170 SNs: 34006BG96 34013BG96 34022BG96 34026BG96 34045BG96 34048BG96 34050BG96 34052BG96 34053BG96 34061BG96 34068BG96 34103BG96 34109BG96 34134BG96 34154BG96 34171BG96 34174BG96 34182BG96 34188BG96 34194BG96 34229BG96 34250BG96 34254BG96 34264BG96 34285BG96 34289BG96 34317BG96 34334BG96 34338BG96 34343BG96 34348BG96 34359BG96 34383BG96 34392BG96 34422BG96 34425BG96 34426BG96 34430BG96 34431BG96 34434BG96 34445BG96 34457BG96 34479BG96 34486BG96 34503BG96 34507BG96 34532BG96 34549BG96 34556BG96 34584BG96 34591BG96 34592BG96 34595BG96 34602BG96 34603BG96 34647BG96 34650GB96…
- Distribution
- Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.
- Recalling firm
- Abbott Laboratories, Abbott Park, IL, United States
- How the firm notified
- Letter
- FDA classified
- Apr 22, 2025
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software…Z-1652-2025 · Class II · Mar 18, 2025 · Ongoing
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software…Z-1654-2025 · Class II · Mar 18, 2025 · Ongoing
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-13 Software Version: All software…Z-1655-2025 · Class II · Mar 18, 2025 · Ongoing
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software…Z-1656-2025 · Class II · Mar 18, 2025 · Ongoing
Newest first, one product per recall event; search them all
- Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20Z-0961-2026 · Class II · Nov 13, 2025 · Ongoing
- Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version:…Z-0489-2026 · Class II · Sep 18, 2025 · Ongoing
- Alinity hq Analyzer, REF: 09P68-01Z-2681-2024 · Class II · Jun 14, 2024 · Ongoing
- ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.Z-1476-2024 · Class II · Feb 21, 2024 · Ongoing
- Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module is a fully automated chemistry analyzer…Z-0681-2024 · Class II · Oct 19, 2023 · Ongoing
- Abbott CELL-DYN Ruby, Model CD-Ruby, List number 08H67-01, and Abbott CELL-DYN Sapphire, Model CD Sapphire, list number…Z-0604-2024 · Class II · Oct 19, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports