Discovery Model Nm/Ct 670

System, Tomography, Computed, Emission. 510(k) K093514, from Ge Medical Systems F.I. Haifa (Tirat Hacarmel, IL). Product code KPS, class 2.

As of Nov 18, 2026, K093514 is an FDA 510(k) held by Ge Medical Systems F.I. Haifa, substantially equivalent on Dec 10, 2009 after 27 days, with 21 recalls tied to it in FDA's recall database.

  • Review time: 27 days
  • Recalls: 21
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K093514, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 18, 2024Z-0894-2025, Z-0891-2025, Z-0893-2025, Z-0895-2025GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could…Open
Dec 19, 2022Z-1053-2023A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.Open
Feb 7, 2021Z-1547-2021GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan…Open
Dec 10, 2020Z-0857-2021GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.Terminated
Aug 14, 2020Z-2998-2020, Z-2996-2020, Z-2997-2020Rotor bearing screws were found loose on detector.Terminated
Mar 20, 2019Z-1259-2019, Z-1260-2019, Z-1265-2019During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design…Terminated
May 24, 2018Z-0608-2019, Z-0607-2019, Z-0606-2019When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off…Completed
May 19, 2017Z-2194-2017It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.Terminated
Oct 25, 2013Z-0367-2014GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640,…Terminated
Jun 13, 2013Z-1792-2013GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE…Terminated
Oct 12, 2012Z-0121-2013GE Healthcare has recently become aware of a potential safety issue associated with the table pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact operation safety. At scan setup, during…Terminated
Jan 11, 2012Z-1110-2012GE Healthcare has recently become aware of a potential safety issue associated with the collimator cart and lead cover of your Discovery NM/CT 670 and/or Discovery NM 630 system that may impact operator safety. 1. The…Terminated

Other KPS clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device

Record

Type
Traditional, third-party review
Received
Nov 13, 2009
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files