Discovery Model Nm/Ct 670
System, Tomography, Computed, Emission. 510(k) K093514, from Ge Medical Systems F.I. Haifa (Tirat Hacarmel, IL). Product code KPS, class 2.
As of Nov 18, 2026, K093514 is an FDA 510(k) held by Ge Medical Systems F.I. Haifa, substantially equivalent on Dec 10, 2009 after 27 days, with 21 recalls tied to it in FDA's recall database.
- Review time: 27 days
- Recalls: 21
- Cited as a predicate by: 5 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Ge Quasar Nuclear Medicine SystemK022960 · Ge Medical Systems F.I. Haifa · Sep 19, 2002
- Ge Brightspeed Delight CT Scanner SystemK082816 · Ge Medical Systems, LLC · Oct 8, 2008
- Symbia, Version 4.0K082506 · Siemens Medical Solutions USA, Inc. · Nov 21, 2008
- Hawkeye 4 Option for Dual-Head Variable Angle Gamma CameraK052434 · GE Medical Systems · Sep 27, 2005
Cited as a predicate
5 later 510(k) summaries name this device as a predicate- AuroraK243605 · Ge Medical Systems Israel, Functional Imaging · Jan 17, 2025
- Nm/Ct 850, Nm/Ct 860K173816 · GE Healthcare · Apr 17, 2018
- Discovery NM/CT 670 CZTK153402 · GE Healthcare · Jan 21, 2016
- Optima Nm/Ct 640K121019 · GE Healthcare · Apr 19, 2012
- Discovery NM 630K111445 · GE Healthcare · Jun 3, 2011
Recalls
Recalls FDA ties to K093514, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Dec 18, 2024 | Z-0894-2025, Z-0891-2025, Z-0893-2025, Z-0895-2025GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could… | Open |
| Dec 19, 2022 | Z-1053-2023A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury. | Open |
| Feb 7, 2021 | Z-1547-2021GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan… | Open |
| Dec 10, 2020 | Z-0857-2021GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector. | Terminated |
| Aug 14, 2020 | Z-2998-2020, Z-2996-2020, Z-2997-2020Rotor bearing screws were found loose on detector. | Terminated |
| Mar 20, 2019 | Z-1259-2019, Z-1260-2019, Z-1265-2019During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design… | Terminated |
| May 24, 2018 | Z-0608-2019, Z-0607-2019, Z-0606-2019When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off… | Completed |
| May 19, 2017 | Z-2194-2017It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed. | Terminated |
| Oct 25, 2013 | Z-0367-2014GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640,… | Terminated |
| Jun 13, 2013 | Z-1792-2013GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE… | Terminated |
| Oct 12, 2012 | Z-0121-2013GE Healthcare has recently become aware of a potential safety issue associated with the table pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact operation safety. At scan setup, during… | Terminated |
| Jan 11, 2012 | Z-1110-2012GE Healthcare has recently become aware of a potential safety issue associated with the collimator cart and lead cover of your Discovery NM/CT 670 and/or Discovery NM 630 system that may impact operator safety. 1. The… | Terminated |
Other KPS clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7 (11364753);Symbia Pro.specta VA40…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI Panvivo (uMI Panvivo ES)K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series Digital SPECT/CT System…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Nov 13, 2009
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files