Discovery NM/CT 670 CZT

System, Tomography, Computed, Emission. 510(k) K153402, from GE Healthcare (Tirat Hacarmel, IL). Product code KPS, class 2.

As of Nov 17, 2026, K153402 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Jan 21, 2016 after 58 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 21, 2016
  • Review time: 58 days
  • Recalls: 6
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The GE Discovery NM/CT 670 CZT system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K153402, newest first; each opens FDA's recall record
Dec 19, 2022Z-1057-2023A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.Open, Classified
Dec 10, 2020Z-0855-2021GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.Terminated
Aug 14, 2020Z-2993-2020Rotor bearing screws were found loose on detector.Terminated
Mar 20, 2019Z-1264-2019During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design…Terminated
Mar 20, 2019Z-1258-2019During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design…Terminated
May 19, 2017Z-2195-2017It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.Terminated

Other KPS Clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device

Record

Type
Traditional
Received
Nov 24, 2015
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files