Nm/Ct 850, Nm/Ct 860

System, Tomography, Computed, Emission. 510(k) K173816, from GE Healthcare (Tirat Hacarmel, IL). Product code KPS, class 2.

As of Nov 18, 2026, K173816 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Apr 17, 2018 after 123 days, with 10 recalls tied to it in FDA's recall database.

  • Review time: 123 days
  • Recalls: 10
  • Cited as a predicate by: 1 device

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • AuroraK243605 · Ge Medical Systems Israel, Functional Imaging · Jan 17, 2025

Recalls

Recalls FDA ties to K173816, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 19, 2022Z-1058-2023A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.Open
Feb 7, 2021Z-1545-2021, Z-1546-2021GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan…Open
Dec 10, 2020Z-0854-2021, Z-0853-2021GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.Terminated
Aug 14, 2020Z-2991-2020, Z-2992-2020Rotor bearing screws were found loose on detector.Terminated
Mar 20, 2019Z-1263-2019, Z-1261-2019, Z-1262-2019During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design…Terminated

Other KPS clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device

Record

Type
Traditional
Received
Dec 15, 2017
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files