Arca
Skip to content

Hawkeye 4 Option For Dual-Head Variable Angle Gamma Camera

System, Tomography, Computed, Emission. 510(k) K052434, from GE Medical Systems (Tirat Hacarmel, IL). Product code KPS, class 2.

As of Nov 17, 2026, K052434 is an FDA 510(k) held by GE Medical Systems, substantially equivalent on Sep 27, 2005 after 21 days, with 8 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 27, 2005
  • Review time: 21 days
  • Recalls: 8
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Intended use of the device is to produce attenuation- corrected NM Images. The attenuation maps are also displayed with the NM images to facilitate the localization of the NM activity in the patient anatomy.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K052434, newest first; each opens FDA's recall record
Oct 14, 2014Z-0187-2015GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems. This check is being added to prevent a…Terminated
Oct 14, 2014Z-0189-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the…Terminated
Oct 14, 2014Z-0184-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the…Terminated
Oct 14, 2014Z-0186-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the…Terminated
Oct 14, 2014Z-0188-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the…Terminated
Oct 14, 2014Z-0185-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the…Terminated
Jun 13, 2013Z-1787-2013GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.Terminated
Mar 6, 2012Z-1078-2013Accelerated fatigue of the lateral axis motion subsystem of the Infinia Computed Emission Tomography System could result in mechanical failure causing the detector to slide; may impact patient and operators.Terminated

Other KPS Clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device

Record

Type
Special
Received
Sep 6, 2005
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.