Hawkeye Option For Dual-Head Variable Angle Gamma Camera
System, Tomography, Computed, Emission. 510(k) K991841, from Ge Medical Systems F.I. Haifa (Tirat Hacarmel, IL). Product code KPS, class 2.
As of Nov 17, 2026, K991841 is an FDA 510(k) held by Ge Medical Systems F.I. Haifa, substantially equivalent on Aug 26, 1999 after 90 days, with 8 recalls tied to it in FDA's recall database.
- Substantially equivalent: Aug 26, 1999
- Review time: 90 days
- Recalls: 8
- Cited as a predicate by: 4 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Vtransact: Attenuation Correction System For Dual-Head Variable Angle Gamma CameraK980959 · Ge Medical Systems F.I. Haifa · Jun 4, 1998
- Advance Analysis SoftwareK941223 · General Electric Co. · Nov 29, 1994
- VaricamK953801 · Elscint, Inc. · Nov 22, 1995
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Hawkeye 4 Option For Dual-Head Variable Angle Gamma CameraK052434 · GE Medical Systems · Sep 27, 2005
- Infinia LightspeedK043381 · Ge Medical Systems F.I. Haifa · Dec 23, 2004
- Siemens Enhanced Imaging SystemK041166 · Siemens Medical Solutions USA, Inc. · May 17, 2004
- Ge Quasar Nuclear Medicine SystemK022960 · Ge Medical Systems F.I. Haifa · Sep 19, 2002
Recalls
Recalls FDA ties to K991841, newest first; each opens FDA's recall record| Oct 14, 2014 | Z-0187-2015GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems. This check is being added to prevent a… | Terminated |
| Oct 14, 2014 | Z-0189-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the… | Terminated |
| Oct 14, 2014 | Z-0184-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the… | Terminated |
| Oct 14, 2014 | Z-0186-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the… | Terminated |
| Oct 14, 2014 | Z-0188-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the… | Terminated |
| Oct 14, 2014 | Z-0185-2015GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the… | Terminated |
| Jun 13, 2013 | Z-1787-2013GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. | Terminated |
| Mar 6, 2012 | Z-1078-2013Accelerated fatigue of the lateral axis motion subsystem of the Infinia Computed Emission Tomography System could result in mechanical failure causing the detector to slide; may impact patient and operators. | Terminated |
Other KPS Clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 28, 1999
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files