Brightview Vct Imaging System

System, Tomography, Computed, Emission. 510(k) K080927, from Philips Medical Systems (Cleveland), Inc. (Milpitas, CA). Product code KPS, class 2.

As of Nov 17, 2026, K080927 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on Apr 11, 2008 after 9 days, with 25 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 11, 2008
  • Review time: 9 days
  • Recalls: 25+
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

K08D927 BrightView VCT Imaging System BrightView VCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with an attenuation device consisting of flat panel x-ray imaging components. BrightView VCT produces non-attenuation corrected SPECT images and attenuation corrected SPECT images with x-ray transmission data that may also be used for scatter correction.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K080927, newest first; the 25 newest; each opens FDA's recall record
Dec 15, 2023Z-0930-2024Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as…Open, Classified
Dec 15, 2023Z-0931-2024Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as…Open, Classified
Nov 30, 2023Z-0679-2024While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients…Open, Classified
Nov 30, 2023Z-0678-2024While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients…Open, Classified
Jun 10, 2019Z-2010-2019An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.Terminated
Jun 10, 2019Z-2011-2019An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.Terminated
Jun 10, 2019Z-2009-2019An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.Terminated
Feb 24, 2017Z-1753-2017The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.Terminated
Feb 24, 2017Z-1750-2017The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.Terminated
Feb 24, 2017Z-1752-2017The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.Terminated
Feb 24, 2017Z-1751-2017The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.Terminated
Jan 31, 2017Z-1481-2017Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes…Terminated
Jan 31, 2017Z-1480-2017Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes…Terminated
Jan 31, 2017Z-1479-2017Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes…Terminated
Dec 23, 2016Z-1186-2017The intended use listed in the English and localized language IFUs for BrightView XCT systems did not match the intended use documented in the 510K submission.Terminated
Dec 23, 2016Z-1187-2017The intended use listed in the English and localized language IFUs for BrightView XCT systems did not match the intended use documented in the 510K submission.Terminated
May 25, 2016Z-2736-2016The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow…Terminated
May 25, 2016Z-2737-2016The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow…Terminated
Dec 17, 2015Z-1022-2016The firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with…Terminated
Mar 6, 2015Z-1792-2015The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or…Terminated
Nov 12, 2014Z-1011-2015Unintended detector and gantry movement due to software issues.Terminated
Nov 4, 2014Z-0450-2015Software issuesTerminated
Jun 23, 2014Z-1982-2014Flat Panel Display (FPD) failed to remain securely locked in the deployed position.Terminated
Dec 19, 2013Z-1074-2014Three problems were reported to the firm: When positioning for a relative 180 degree non circular acquisition the second (bottom) detector was not positioned correctly causing an inappropriate rotation orbit…Terminated
Oct 23, 2013Z-0203-2015It was discovered the CT portion of the scan was interrupted ( stopping the CT exposure at the time of interruption) and did not complete successfully.Terminated

Other KPS Clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device

Record

Type
Traditional, third-party review
Received
Apr 2, 2008
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files