Apollo Gamma Camera System
System, Tomography, Computed, Emission. 510(k) K062298, from Adac Laboratories (Milpitas, CA). Product code KPS, class 2.
As of Nov 18, 2026, K062298 is an FDA 510(k) held by Adac Laboratories, substantially equivalent on Aug 23, 2006 after 15 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 15 days
- Recalls: 25+
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Apollo Gamma Camera System is intended to produce images depicting the anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Forte Gamma Camera SystemK033254 · Adac Laboratories · Oct 22, 2003
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Brightview Vct Imaging SystemK080927 · Philips Medical Systems (Cleveland), Inc. · Apr 11, 2008
Recalls
Recalls FDA ties to K062298, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Dec 15, 2023 | Z-0929-2024Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as… | Open |
| Nov 30, 2023 | Z-0677-2024While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients… | Open |
| Dec 15, 2017 | Z-0511-2018, Z-0512-2018, Z-0510-2018During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially… | Terminated |
| Feb 24, 2017 | Z-1753-2017, Z-1750-2017, Z-1752-2017, Z-1751-2017The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive. | Terminated |
| Jan 31, 2017 | Z-1481-2017, Z-1480-2017, Z-1479-2017Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes… | Terminated |
| May 25, 2016 | Z-2735-2016The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow… | Terminated |
| Dec 17, 2015 | Z-1020-2016, Z-1021-2016The firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with… | Terminated |
| Mar 6, 2015 | Z-1791-2015, Z-1790-2015The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or… | Terminated |
| Nov 4, 2014 | Z-0450-2015Software issues | Terminated |
| Apr 25, 2013 | Z-1481-2013Philips has received one report from the field that there was an unexpected motion of the detector due to a failure of the mechanical assembly. | Terminated |
| Jun 15, 2011 | Z-2947-2011Non-conformances with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly. | Terminated |
| Jun 3, 2011 | Z-2867-2011The gap between the two parts of the table may permit pinches to occur on either patients skin or other body parts; it has the potential for serious injury if a patients finger is caught between the pallet and patient… | Terminated |
| Jun 3, 2011 | Z-2868-2011, Z-2870-2011, Z-2869-2011, Z-2872-2011The gap between the two parts of the table may permit pinches to occur on either patient's skin or other body parts; it has the potential for serious injury if a patient's finger is caught between the pallet and patient… | Terminated |
Other KPS clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7 (11364753);Symbia Pro.specta VA40…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI Panvivo (uMI Panvivo ES)K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series Digital SPECT/CT System…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Aug 8, 2006
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files