Ge Brightspeed Delight CT Scanner System

System, X-Ray, Tomography, Computed. 510(k) K082816, from Ge Medical Systems, LLC (Waukesha, WI). Product code JAK, class 2.

As of Nov 18, 2026, K082816 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Oct 8, 2008 after 13 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 13 days
  • Recalls: 9
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The GE BrightSpeed Delight Series Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, patient for all ages, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K082816, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 24, 2018Z-2586-2018Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).Terminated
Apr 27, 2018Z-2872-2018, Z-2873-2018GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.Terminated
May 19, 2017Z-2188-2017, Z-2186-2017It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.Terminated
Oct 9, 2013Z-0840-2014There is an issue with the Manual Film Composer feature on some CT products. There is an opportunity, while following a specific workflow, to create a film image with one patient's images and another patient's name in…Terminated
Jan 19, 2011Z-2214-2011GE Healthcare has become aware of a potential set of circumstances that could cause X-ray continuation during an unexpected table stop on certain BrightSpeed systems (BrightSpeed Excel/Edge/Elite Select and BrightSpeed…Terminated
Sep 3, 2010Z-0368-2011, Z-0369-2011GE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems.Terminated

Other JAK clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional, third-party review
Received
Sep 25, 2008
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files