Arca
Skip to content

Symbia, Version 4.0

System, Tomography, Computed, Emission. 510(k) K082506, from Siemens Medical Solutions USA, Inc. (Hoffman Estates, IL). Product code KPS, class 2.

As of Nov 17, 2026, K082506 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Nov 21, 2008 after 84 days, with 9 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 21, 2008
  • Review time: 84 days
  • Recalls: 9
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K082506, newest first; each opens FDA's recall record
Jul 28, 2014Z-2713-2014Siemens Medical Solutions USA, Inc. is performing a field correction because we recently detected a rubber isolator with unfavorable aging properties. The function of the rubber isolator, which is situated within the…Terminated
Jul 28, 2014Z-2712-2014Siemens Medical Solutions USA, Inc. is performing a field correction because we recently detected a rubber isolator with unfavorable aging properties. The function of the rubber isolator, which is situated within the…Terminated
Mar 5, 2014Z-0960-2015There are confirmed reports of a component coming loose inside the CT enclosure and damaging the system. components.Terminated
Oct 28, 2013Z-0115-2015Reports of loose bolts in the detector support structure of Symbia and E.CAM systems.Terminated
Aug 16, 2013Z-0062-2017The detectors on the Symbia S and Symbia T camera systems are supported and moved horizontally and vertically by four linear bearing cars. There is a potential for the linear bearing car to fail allowing the ball…Terminated
Oct 23, 2012Z-1087-2015There is a potential for the linear bearing car to fail allowing the ball bearings to fall out.Terminated
Oct 14, 2011Z-0283-2012Errors could potentially impact image interpretation. The acquisition software is exhibiting a defect in the energy window position when an isotope is peaked. There is a potential patient safety issue when running syngo…Terminated
Oct 14, 2011Z-0284-2012Errors could potentially impact image interpretation. The acquisition software is exhibiting a defect in the energy window position when an isotope is peaked. There is a potential patient safety issue when running syngo…Terminated
Jun 22, 2011Z-2980-2011There a potential patient safety issue when using the Symbia Radiation Therapy Pallet. When a patient is not iso-centered before starting a scan with the Symbia T Series Systems and a non-circular orbit is selected, it…Terminated

Other KPS Clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device

Record

Type
Traditional
Received
Aug 29, 2008
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.