Symbia, Version 4.0
System, Tomography, Computed, Emission. 510(k) K082506, from Siemens Medical Solutions USA, Inc. (Hoffman Estates, IL). Product code KPS, class 2.
As of Nov 17, 2026, K082506 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Nov 21, 2008 after 84 days, with 9 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 21, 2008
- Review time: 84 days
- Recalls: 9
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Siemens Enhanced Imaging SystemK041166 · Siemens Medical Solutions USA, Inc. · May 17, 2004
- Symbia-E, Model 10275879K072567 · Siemens Medical Solutions USA, Inc. · Oct 30, 2007
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- SymbiaK131634 · Siemens Medical Solutions USA, Inc. · Aug 16, 2013
- Symbia.NetK110844 · Siemens Medical Solutions USA, Inc. · Apr 20, 2011
- Symbia.Net, Symbia.Net Select Model 10520828, 10520827K100619 · Siemens Medical Solutions USA, Inc. · May 7, 2010
- Discovery Model Nm/Ct 670K093514 · Ge Medical Systems F.I. Haifa · Dec 10, 2009
Recalls
Recalls FDA ties to K082506, newest first; each opens FDA's recall record| Jul 28, 2014 | Z-2713-2014Siemens Medical Solutions USA, Inc. is performing a field correction because we recently detected a rubber isolator with unfavorable aging properties. The function of the rubber isolator, which is situated within the… | Terminated |
| Jul 28, 2014 | Z-2712-2014Siemens Medical Solutions USA, Inc. is performing a field correction because we recently detected a rubber isolator with unfavorable aging properties. The function of the rubber isolator, which is situated within the… | Terminated |
| Mar 5, 2014 | Z-0960-2015There are confirmed reports of a component coming loose inside the CT enclosure and damaging the system. components. | Terminated |
| Oct 28, 2013 | Z-0115-2015Reports of loose bolts in the detector support structure of Symbia and E.CAM systems. | Terminated |
| Aug 16, 2013 | Z-0062-2017The detectors on the Symbia S and Symbia T camera systems are supported and moved horizontally and vertically by four linear bearing cars. There is a potential for the linear bearing car to fail allowing the ball… | Terminated |
| Oct 23, 2012 | Z-1087-2015There is a potential for the linear bearing car to fail allowing the ball bearings to fall out. | Terminated |
| Oct 14, 2011 | Z-0283-2012Errors could potentially impact image interpretation. The acquisition software is exhibiting a defect in the energy window position when an isotope is peaked. There is a potential patient safety issue when running syngo… | Terminated |
| Oct 14, 2011 | Z-0284-2012Errors could potentially impact image interpretation. The acquisition software is exhibiting a defect in the energy window position when an isotope is peaked. There is a potential patient safety issue when running syngo… | Terminated |
| Jun 22, 2011 | Z-2980-2011There a potential patient safety issue when using the Symbia Radiation Therapy Pallet. When a patient is not iso-centered before starting a scan with the Symbia T Series Systems and a non-circular orbit is selected, it… | Terminated |
Other KPS Clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 29, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files