Intuitive Surgical Endoscopic Instrument Control Systems, Models IS1200, IS2000 And IS3000

System, Surgical, Computer Controlled Instrument. 510(k) K090993, from Intuitive Surgical, Inc. (Sunnyvale, CA). Product code NAY, class 2.

As of Nov 17, 2026, K090993 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on Dec 16, 2009 after 253 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 16, 2009
  • Review time: 253 days
  • Recalls: 3
  • Cited as a predicate by: 20 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Intuitive Surgical Endoscopic Instrument Control Systems (da Vinci, da Vinci S and da Vinci Si Surgical Systems Models IS1200, IS2000, IS3000) are intended to assist in the accurate control of Intuitive Surgical Endoscopic EndoWrist Instruments and Accessories including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic/harmonic shears, forceps/pick-ups…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

20 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K090993, newest first; each opens FDA's recall record
Jan 24, 2013Z-1424-2013Change to User Manual to remove one indication for use and amend warnings to emphasize restrictions for use.Terminated
Oct 13, 2011Z-1426-2013Promotional literature for use of the da Vinci system for thyroidectomy, which had not been approved with a change to the 510(k).Terminated
Jul 5, 2011Z-2930-2011Potential failure of the retention component of the Master Tool Manipulator (MTM) of the Intuitive Surgical da Vinci Si Surgical System which could cause uncontrolled movement.Terminated

Other NAY Clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device

Record

Type
Traditional
Received
Apr 7, 2009
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.