Intuitive Surgical Da Vinci Surgical System, Model IS2000
System, Surgical, Computer Controlled Instrument. 510(k) K050369, from Intuitive Surgical, Inc. (Sunnyvale, CA). Product code NAY, class 2.
As of Nov 17, 2026, K050369 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on Apr 29, 2005 after 74 days, with 25 recalls tied to it in FDA's recall database.
- Substantially equivalent: Apr 29, 2005
- Review time: 74 days
- Recalls: 25+
- Cited as a predicate by: 22 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Intuitive Surgical (6) da Vinci (6) Surgical System is intended to assist in the accurate control of Intuitive Surgical (6) Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Intuitive Surgical Endoscopic Instruments, Intuitive Surgical Endoscopic Instrument…K990144 · Intuitive Surgical, Inc. · Jul 11, 2000
- Intuitive Surgical Da Vinci Endoscopic Control SystemK002489 · Intuitive Surgical, Inc. · Mar 2, 2001
- Intuitive Surgical Da Vinci Surgical System, Model Isi 1000K011002 · Intuitive Surgical, Inc. · May 30, 2001
- Intuitive Surgical Endowrist Endoscopic Instrument FamilyK013416 · Intuitive Surgical, Inc. · Jan 10, 2002
- Intuitive Surgical Da Vinci Surgical System, Model IS1000K021036 · Intuitive Surgical, Inc. · Jun 26, 2002
- Intuitive Surgical Endoscopic Instrument Control System & Endoscopic Instruments, Model…K022574 · Intuitive Surgical, Inc. · Nov 12, 2002
- Intuitive Surgical Da Vinci Endoscopic Instrument Control System And Endoscopic…K040237 · Intuitive Surgical, Inc. · Jul 7, 2004
- Intuitive Surgical Da Vinci Surgical System And Endoscopic Instruments, Models IS1200 &…K043153 · Intuitive Surgical, Inc. · Dec 15, 2004
- Intuitive Surgical Monopolar Curved Scissors, Model 400179; Tip Cover Accessory, Model…K050005 · Intuitive Surgical, Inc. · Jan 25, 2005
Cited as a Predicate
22 later 510(k) summaries name this device as a predicate; the 12 newest- Remanufactured EndoWrist Tenaculum Forceps (420207)K250539 · Rebotix · Aug 19, 2025
- Remanufactured EndoWrist Cobra Grasper (420190)K250417 · Rebotix · Aug 20, 2025
- Remanufactured EndoWrist Cadiere Forceps (420049)K250399 · Rebotix · Aug 19, 2025
- Remanufactured EndoWrist Long Tip Forceps (420048)K250387 · Rebotix · Aug 19, 2025
- Remanufactured EndoWrist ProGrasp Forceps (420093)K241872 · Rebotix · Nov 7, 2024
- 8mm Monopolar Curved ScissorsK210478 · Iconocare Health · Sep 30, 2022
- ArtiSential Bipolar Fenestrated ForcepsK190909 · Livsmed, Inc. · Feb 13, 2020
- 8mm Monopolar Curved ScissorsK180033 · Intuitive Surgical, Inc. · Apr 6, 2018
- EndoWrist 5mm Thoracic GrasperK173415 · Intuitive Surgical, Inc. · May 31, 2018
- da Vinci S/Si Endoscopes, da Vinci Xi EndoscopesK170641 · Intuitive Surgical, Inc. · Sep 21, 2017
- Da Vinci Sp Surgical System, Endowrist Sp Instruments, And AccessoriesK131962 · Intuitive Surgical, Inc. · Apr 17, 2014
- Da Vinvi Surgical SystemK123329 · Intuitive Surgical, Inc. · Sep 17, 2014
Recalls
Recalls FDA ties to K050369, newest first; the 25 newest; each opens FDA's recall record| Dec 9, 2025 | Z-1696-2026Due to increased complaints for broken/frayed grip cables for reusable instruments. | Open, Classified |
| Dec 9, 2025 | Z-1700-2026Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments | Open, Classified |
| Dec 9, 2025 | Z-1698-2026Due to increased complaints for broken/frayed grip cables for reusable instruments. | Open, Classified |
| Dec 9, 2025 | Z-1699-2026Due to increased complaints for broken/frayed grip cables for reusable instruments. | Open, Classified |
| Dec 9, 2025 | Z-1701-2026Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments | Open, Classified |
| Dec 9, 2025 | Z-1697-2026Due to increased complaints for broken/frayed grip cables for reusable instruments. | Open, Classified |
| Jul 1, 2022 | Z-1459-2022Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the system. | Open, Classified |
| Aug 15, 2018 | Z-3024-2018Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of the extension tube. These cracks may fracture the plastic surrounding the instrument shaft, and may cause… | Open, Classified |
| Mar 16, 2016 | Z-1376-2016Intuitive Surgical has found that the use of instruments other than the Cardiac Probe Grasper (e.g. Large Needle Driver Instrument) to manipulate the ablation probes during cardiac surgery can result in an increase of… | Terminated |
| May 21, 2015 | Z-1812-2015The coating of the 5 mm Thoracic Grasper main tube may experience damage with the possibility of fragments of coating material dislodging and potentially falling into patient. | Terminated |
| Jun 13, 2014 | Z-2020-2014Manufacturing variation in the grip assembly of the Large Clip applier which can lead to the instrument jamming during insertion when used with the 8 mm cannula or with certain reducer accessories.. | Terminated |
| May 21, 2014 | Z-1966-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1968-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1973-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1967-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1965-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1974-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1972-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1975-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1964-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1969-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1970-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| May 21, 2014 | Z-1971-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the… | Terminated |
| Nov 15, 2013 | Z-0520-2014Firm has become aware of the potential for detachment of the jaw insert for Large Needle Drivers and Mega Needle Drivers. | Terminated |
| Nov 15, 2013 | Z-0521-2014Firm has become aware of the potential for detachment of the jaw insert for Large Needle Drivers and Mega Needle Drivers. | Terminated |
Other NAY Clearances
The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device- da Vinci Xi Surgical System (IS4000);da Vinci X Surgical System (IS4200)K263592 · Intuitive Surgical, Inc. · Oct 30, 2026
- da Vinci Surgical System, Model IS5000K261823 · Intuitive Surgical, Inc. · Sep 8, 2026
- da Vinci Surgical System, Model IS5000K261824 · Intuitive Surgical, Inc. · Jul 31, 2026
- da Vinci Firefly Imaging SystemK260695 · Intuitive Surgical · May 22, 2026
- Senhance Ultrasonic SystemK254192 · Asensus Surgical · May 12, 2026
- da Vinci Force Feedback InstrumentsK253986 · Intuitive Surgical, Inc. · Mar 11, 2026
- da Vinci Surgical System (IS5000)K251739 · Intuitive Surgical, Inc. · Jan 16, 2026
- da Vinci SP Surgical System (SP1098); da Vinci E-200 GeneratorK253556 · Intuitive Surgical, Inc. · Jan 16, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 14, 2005
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files