System, Surgical, Computer Controlled Instrument. 510(k) K990144, from Intuitive Surgical, Inc. (Mountain View, CA). Product code NAY, class 2.

As of Nov 18, 2026, K990144 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on Jul 11, 2000 after 539 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 539 days
  • Recalls: 3
  • Cited as a predicate by: 21 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Intuitive Surgical™ Endoscopic Instrument Control System (hereinafter referred to as the “da Vinci™ System”) is intended to assist in the accurate control of Intuitive Surgical™ endoscopic instruments including: rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps / pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

21 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K990144, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 4, 2007Z-0152-2008Under-Rated Fuses -- The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this…Terminated
Oct 4, 2007Z-0151-2008Under-Rated Fuses--The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this…Terminated
May 3, 2007Z-1244-2007Under certain circumstances, the product's software may crash and require a manual override or restart before functioning again. In addition, there is an unrelated addendum to the User Manual to clarify an icon not…Terminated

Other NAY clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device

Record

Type
Traditional
Received
Jan 19, 1999
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files