Arca
Skip to content

Da Vinci S Surgical SYSTEM-V1.1, Model IS2000

System, Surgical, Computer Controlled Instrument. 510(k) K063220, from Intuitive Surgical, Inc. (Sunnyvale, CA). Product code NAY, class 2.

As of Nov 17, 2026, K063220 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on Dec 1, 2006 after 38 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 1, 2006
  • Review time: 38 days
  • Recalls: 4
  • Cited as a predicate by: 7 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K063220, newest first; each opens FDA's recall record
Feb 6, 2012Z-1202-2012The holding brake, a component of the da Vinci S Surgical System, may allow passive uncontrolled motion due to gravity during specific power-off conditions.Terminated
Oct 4, 2007Z-0152-2008Under-Rated Fuses -- The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this…Terminated
Oct 4, 2007Z-0151-2008Under-Rated Fuses--The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this…Terminated
May 14, 2007Z-1245-2007Product may malfunction and fail to start up on AC power.Terminated

Other NAY Clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device

Record

Type
Traditional
Received
Oct 24, 2006
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.