Intuitive Surgical Da Vinci Si Surgical System: Model IS3000

System, Surgical, Computer Controlled Instrument. 510(k) K081137, from Intuitive Surgical, Inc. (Sunnyvale, CA). Product code NAY, class 2.

As of Nov 20, 2026, K081137 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on Feb 18, 2009 after 303 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 303 days
  • Recalls: 25+
  • Cited as a predicate by: 33 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Intuitive Surgical Endoscopic Instrument Control System (Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

33 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K081137, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 16, 2018Z-3254-2018Certain Redundant Medical Grade Power Supply (RMGPS) units specific to Surgeon Console and Patient Side Carts were manufactured incorrectly and may fail prior or during use.Terminated
Mar 16, 2016Z-1377-2016Intuitive Surgical has found that the use of instruments other than the Cardiac Probe Grasper (e.g. Large Needle Driver Instrument) to manipulate the ablation probes during cardiac surgery can result in an increase of…Terminated
Feb 16, 2016Z-1191-2016Terminated
Dec 5, 2014Z-0940-2015Intuitive Surgical is initiating a voluntary correction related to da Vinci Si System (IS3000) Vision Side Carts, with threaded-in caster wheels, which can result in the caster wheel bolts gradually unthreading and the…Terminated
May 21, 2014Z-1966-2014, Z-1968-2014, Z-1973-2014, Z-1967-2014, Z-1965-2014, Z-1974-2014, Z-1972-2014, Z-1975-2014, Z-1964-2014, Z-1969-2014, Z-1970-2014, Z-1971-2014Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the…Terminated
Nov 11, 2013Z-0436-2014Reports of friction within certain instrument arms can interrupt smooth instrument motion. This can be felt by the surgeon as resistance in the movement of the master.In this situation, the instrument can stall…Terminated
Oct 21, 2013Z-0146-2014Intuitive has identified that the metalized coating on the IS3000 HD lamp module may be incompatible with the illuminator control board, leading to an erroneous red Preventive Maintenance Message that shows up on…Terminated
Jun 8, 2013Z-1687-2013, Z-1686-2013Factory testing on da Vinci Si Surgical Systems may not be in compliance with UL Standards as one of the testing devices was found to be working incorrectly.Terminated
Mar 15, 2013Z-1110-2013Revised instructions for users of the da Vinci S, Si and Si-e systems with additional information regarding sterile adapter engagement and instrument installation.Terminated
Apr 27, 2012Z-1994-2014Patient Side Cart battery boxes may overheat in rare charging conditions, causing battery bulging.Terminated
Feb 6, 2012Z-1202-2012The holding brake, a component of the da Vinci S Surgical System, may allow passive uncontrolled motion due to gravity during specific power-off conditions.Terminated
Oct 17, 2011Z-0088-2014Improper restraints during transportation of the da Vinci system could cause the HRSV Monitor to become loose from its mount.Terminated
Nov 30, 12Z-0139-2014Increased number of complaints regarding fluid entering the patient card base of the da Vinci S and SI systems.Terminated

Other NAY clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device

Record

Type
Traditional
Received
Apr 21, 2008
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files