Ge 1.5T And 3.0T Signa Hdx Mr System

System, Nuclear Magnetic Resonance Imaging. 510(k) K052293, from Ge Medical Systems, LLC (Milwaukee, WI). Product code LNH, class 2.

As of Nov 18, 2026, K052293 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Sep 21, 2005 after 29 days, with 21 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 21, 2005
  • Review time: 29 days
  • Recalls: 21
  • Cited as a predicate by: 8 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The GE Signa® HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® HDx MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K052293, newest first; each opens FDA's recall record
Jun 14, 2024Z-2471-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE…Open, Classified
Jun 14, 2024Z-2463-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE…Open, Classified
Feb 18, 2022Z-0860-2022There is potential for the images to be flipped left to right.Open, Classified
Dec 23, 2021Z-0603-2022The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.Open, Classified
Aug 26, 2021Z-0141-2022Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.Terminated
Jun 4, 2021Z-2065-2021GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes…Open, Classified
Dec 23, 2019Z-0877-2020It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This…Terminated
Jun 7, 2018Z-0111-2019The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.Terminated
Jun 7, 2018Z-0112-2019The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.Terminated
Jun 7, 2018Z-0110-2019The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.Terminated
May 11, 2017Z-3027-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.Terminated
Aug 31, 2015Z-0259-2016When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR…Terminated
Aug 31, 2015Z-0258-2016When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR…Terminated
Jun 8, 2012Z-1207-2013GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO…Terminated
Nov 21, 2011Z-1943-2013Screen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the…Terminated
May 10, 2011Z-0935-2013An error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports.Terminated
Nov 9, 2010Z-1428-2011GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the…Terminated
Oct 6, 2009Z-0592-2010GE Healthcare has recently become aware that the use of Confirma, Inc.'s CADstream in conjunction with the GE Healthcare's PURE (Phased Array Uniformity Enhancement) software for dynamic MRI imaging (Signa Excite 1.5T,…Terminated
Mar 28, 2008Z-2225-2008Possible misdiagnosis or inappropriate treatment based on the image. Signal homogeneity may be affected when BRAVO applications are set with IR Prep off. The concern is that signal inhomogeneity can result in an…Terminated
Dec 14, 2007Z-1890-2008GE healthcare has become aware of an artifact which could affect the diagnostic capability associated with the 3.0T Propeller T2 Flair application of the 3.0T Signa HDx scanner.Terminated
Oct 1, 2007Z-2210-2008A software issue that may result in misregistration of functional and anatomical images associated with the 3.0T BrainwaveRT fMRI Application on the 3.0T Signa HDx scanner.Terminated

Other LNH Clearances

The newest 510(k)s for the same product code; their median review was 152 days; see every LNH device

Record

Type
Special
Received
Aug 23, 2005
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files