Ge 1.5T And 3.0T Signa Hdx Mr System
System, Nuclear Magnetic Resonance Imaging. 510(k) K052293, from Ge Medical Systems, LLC (Milwaukee, WI). Product code LNH, class 2.
As of Nov 18, 2026, K052293 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Sep 21, 2005 after 29 days, with 21 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 21, 2005
- Review time: 29 days
- Recalls: 21
- Cited as a predicate by: 8 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe GE Signa® HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® HDx MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Ge Signa Excite 1.5T Mr System, Ge Signa Excite 3.0T Mr SystemK041476 · Ge Medical Systems, LLC · Jun 17, 2004
Cited as a Predicate
8 later 510(k) summaries name this device as a predicate- 1.5T Signa Hdxt 3.0T Signa Hdxt Signa VibrantK121676 · Ge Medical Systems, LLC · Sep 20, 2012
- Optima MR450WK091536 · Ge Medical Systems, LLC · Jul 17, 2009
- Mr-Touch OptionK083421 · Ge Medical Systems, LLC · Jul 24, 2009
- Discovery MR450K083147 · Ge Medical Systems, LLC · Nov 4, 2008
- Ge Signa Mr Oncology PackageK081916 · Ge Medical Systems, LLC · Jul 22, 2008
- Signa 3.0T MR750 SystemK081028 · Ge Medical Systems, LLC · Apr 25, 2008
- Ideal Software Option For Ge Mr ScannersK072998 · Ge Medical Systems, LLC · Nov 1, 2007
- Ge 1.5T Signa Hde Mr SystemK052978 · Ge Medical Systems, LLC · Nov 9, 2005
Recalls
Recalls FDA ties to K052293, newest first; each opens FDA's recall record| Jun 14, 2024 | Z-2471-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE… | Open, Classified |
| Jun 14, 2024 | Z-2463-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE… | Open, Classified |
| Feb 18, 2022 | Z-0860-2022There is potential for the images to be flipped left to right. | Open, Classified |
| Dec 23, 2021 | Z-0603-2022The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements. | Open, Classified |
| Aug 26, 2021 | Z-0141-2022Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images. | Terminated |
| Jun 4, 2021 | Z-2065-2021GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes… | Open, Classified |
| Dec 23, 2019 | Z-0877-2020It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This… | Terminated |
| Jun 7, 2018 | Z-0111-2019The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch. | Terminated |
| Jun 7, 2018 | Z-0112-2019The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch. | Terminated |
| Jun 7, 2018 | Z-0110-2019The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch. | Terminated |
| May 11, 2017 | Z-3027-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard. | Terminated |
| Aug 31, 2015 | Z-0259-2016When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR… | Terminated |
| Aug 31, 2015 | Z-0258-2016When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR… | Terminated |
| Jun 8, 2012 | Z-1207-2013GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO… | Terminated |
| Nov 21, 2011 | Z-1943-2013Screen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the… | Terminated |
| May 10, 2011 | Z-0935-2013An error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports. | Terminated |
| Nov 9, 2010 | Z-1428-2011GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the… | Terminated |
| Oct 6, 2009 | Z-0592-2010GE Healthcare has recently become aware that the use of Confirma, Inc.'s CADstream in conjunction with the GE Healthcare's PURE (Phased Array Uniformity Enhancement) software for dynamic MRI imaging (Signa Excite 1.5T,… | Terminated |
| Mar 28, 2008 | Z-2225-2008Possible misdiagnosis or inappropriate treatment based on the image. Signal homogeneity may be affected when BRAVO applications are set with IR Prep off. The concern is that signal inhomogeneity can result in an… | Terminated |
| Dec 14, 2007 | Z-1890-2008GE healthcare has become aware of an artifact which could affect the diagnostic capability associated with the 3.0T Propeller T2 Flair application of the 3.0T Signa HDx scanner. | Terminated |
| Oct 1, 2007 | Z-2210-2008A software issue that may result in misregistration of functional and anatomical images associated with the 3.0T BrainwaveRT fMRI Application on the 3.0T Signa HDx scanner. | Terminated |
Other LNH Clearances
The newest 510(k)s for the same product code; their median review was 152 days; see every LNH device- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- MagicScan-1.5, GreenMR Serenity 1.5TK254250 · Xingaoyi Medical Equipment Co.,Ltd · Aug 17, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Aug 23, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files