Optima MR450W
System, Nuclear Magnetic Resonance Imaging. 510(k) K091536, from Ge Medical Systems, LLC (Waukesha, WI). Product code LNH, class 2.
As of Nov 19, 2026, K091536 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Jul 17, 2009 after 52 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 52 days
- Recalls: 8
- Cited as a predicate by: 3 devices
The Optima MR450w is a whole body magnetic resonance scanner designed to support high resolution and high signal-to-noise ratio images in short exam times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique anatomical images, spectroscopic data, parametric maps, or dynamic images of the structures or functions of the entire body.
- Discovery MR450K083147 · Ge Medical Systems, LLC · Nov 4, 2008
- GE 1.5T and 3.0T Signa Hdx Mr SystemK052293 · Ge Medical Systems, LLC · Sep 21, 2005
- Optima MR450WK113490 · Ge Medical Systems, LLC · Jan 13, 2012
- Gem Option for 1.5 Mri SystemsK103335 · GE Healthcare · Mar 7, 2011
- Discovery MR750W 3.0TK103327 · Ge Healthcare Japan Corporation · Sep 30, 2011
| Sep 8, 2020 | Z-0128-2021Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet. | Terminated |
| May 11, 2017 | Z-3024-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard. | Terminated |
| Apr 23, 2014 | Z-1533-2014GE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products. Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block… | Terminated |
| Apr 2, 2014 | Z-1480-2014Potential safety issue with gradient coil electromechanical connections associated with GE MR Products. Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance… | Terminated |
| Jun 8, 2012 | Z-1207-2013GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO… | Terminated |
| Jul 29, 2011 | Z-3121-2011GE Healthcare has become aware of two potential safety issues associated with MR products (Optima MR450w, Discovery MR450, Discovery MR750, Brivo MR355, Optima MR360) that may impact patient safety. 1) Inline… | Terminated |
| May 10, 2011 | Z-0935-2013An error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports. | Terminated |
| Nov 18, 2010 | Z-1377-2011GE Healthcare has become aware that during certain service procedures, there is a potential for electric shock associated with the gradient cooling manifold hoses of your MR750, MR450, and MR450w MR systems that may… | Terminated |
- MAGNETOM Free.XLK262011 · Siemens Healthineers AG · Nov 5, 2026
- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- May 26, 2009
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files