Gem Option for 1.5 Mri Systems
Coil, Magnetic Resonance, Specialty. 510(k) K103335, from GE Healthcare (Aurora, OH). Product code MOS, class 2.
As of Nov 19, 2026, K103335 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Mar 7, 2011 after 112 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 112 days
- Recalls: 3
The GEM Option for 1.5T MRI systems is a set of receive-only RF surface coils designed for use with 1.5T MRI systems manufactured by GE Healthcare. The GEM Option for 1.5T MRI systems is indicated for use for: head, neck, brachial-plexus, spine, pelvis, hips, prostate, abdominal, cardiac, lower extremities, blood vessels, and long bone imaging. The nucleus detected is hydrogen.
- Optima MR450WK091536 · Ge Medical Systems, LLC · Jul 17, 2009
| Apr 23, 2014 | Z-1533-2014GE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products. Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block… | Terminated |
| Apr 2, 2014 | Z-1480-2014Potential safety issue with gradient coil electromechanical connections associated with GE MR Products. Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance… | Terminated |
| Apr 12, 2012 | Z-1492-2012GE Healthcare is conducting a recall on Optima MR450W with GEM. During scanning of a patient, if an improperly torqued gradient connection exists, it may cause the gradient lead to produce excessive heat. If a second… | Terminated |
- InkSpace Imaging Small Body Array (SBA12PH15x)K262813 · Inkspace Imaging, Inc. · Oct 5, 2026
- dS Breast 7ch 1.5TdS Breast 7ch 3.0TdS Breast 16ch 1.5TdS Breast 16ch 3.0TdS Breast Coil 7ch 1.5TdS Sentinelle Breast…K262330 · Philips Medical Systems Nederland B.V. · Jul 30, 2026
- dS FootAnkle 16Ch 1.5T;dS FootAnkle 16Ch 3.0T;dS HiRes Hand/Wrist 16Ch 1.5T;dS HiRes Hand/Wrist 16CH 3.0T;dS Small…K262294 · Philips Medical Systems Nederland B.V. · Jul 24, 2026
- 0.55T 12Ch Breast CoilK260947 · Neocoil, LLC · Apr 23, 2026
- HD 8Ch Wrist ArrayK252587 · Shenzhen RF Tech Co., Ltd. · Apr 14, 2026
- GEM Flex Coil 16-L Array, 3.0T Receive Only; GEM Flex Coil 16-M Array, 3.0T Receive Only; GEM Flex Coil 16-S Array,…K252876 · Shenzhen RF Tech Co., Ltd. · Apr 1, 2026
- Smart Fit TorsoCardiac 1.5TK260519 · Philips Healthcare (Suzhou) Co., Ltd. · Mar 20, 2026
- GEM Flex Coil 16-L Array, 1.5T Receive Only; GEM Flex Coil 16-M Array, 1.5T Receive Only; GEM Flex Coil 16-S Array,…K252179 · Shenzhen RF Tech Co., Ltd. · Feb 10, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 15, 2010
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files