Arca
Skip to content
Tools
Tools

Gem Option for 1.5 Mri Systems

Coil, Magnetic Resonance, Specialty. 510(k) K103335, from GE Healthcare (Aurora, OH). Product code MOS, class 2.

As of Nov 19, 2026, K103335 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Mar 7, 2011 after 112 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 112 days
  • Recalls: 3

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The GEM Option for 1.5T MRI systems is a set of receive-only RF surface coils designed for use with 1.5T MRI systems manufactured by GE Healthcare. The GEM Option for 1.5T MRI systems is indicated for use for: head, neck, brachial-plexus, spine, pelvis, hips, prostate, abdominal, cardiac, lower extremities, blood vessels, and long bone imaging. The nucleus detected is hydrogen.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K103335, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 23, 2014Z-1533-2014GE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products. Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block…Terminated
Apr 2, 2014Z-1480-2014Potential safety issue with gradient coil electromechanical connections associated with GE MR Products. Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance…Terminated
Apr 12, 2012Z-1492-2012GE Healthcare is conducting a recall on Optima MR450W with GEM. During scanning of a patient, if an improperly torqued gradient connection exists, it may cause the gradient lead to produce excessive heat. If a second…Terminated

Other MOS clearances

The newest 510(k)s for the same product code; their median review was 59 days; see every MOS device

Record

Type
Traditional
Received
Nov 15, 2010
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files