Discovery MR450
System, Nuclear Magnetic Resonance Imaging. 510(k) K083147, from Ge Medical Systems, LLC (Waukesha, WI). Product code LNH, class 2.
As of Nov 18, 2026, K083147 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Nov 4, 2008 after 11 days, with 10 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 4, 2008
- Review time: 11 days
- Recalls: 10
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Discovery MR450 is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Ge 1.5T And 3.0T Signa Hdx Mr SystemK052293 · Ge Medical Systems, LLC · Sep 21, 2005
- Signa 3.0T MR750 SystemK081028 · Ge Medical Systems, LLC · Apr 25, 2008
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- SwipK131241 · Philips Medical Systems Nederland B.V. · Aug 30, 2013
- Brivo MR355, Optima MR360K103330 · Ge Hangwei Medical Systems Co., Ltd. · Feb 11, 2011
- 3D AslK092925 · Ge Medical Systems, LLC · Jan 6, 2010
- Optima MR450WK091536 · Ge Medical Systems, LLC · Jul 17, 2009
Recalls
Recalls FDA ties to K083147, newest first; each opens FDA's recall record| May 11, 2017 | Z-3022-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard. | Terminated |
| May 11, 2017 | Z-3023-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard. | Terminated |
| Apr 23, 2014 | Z-1533-2014GE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products. Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block… | Terminated |
| Apr 2, 2014 | Z-1480-2014Potential safety issue with gradient coil electromechanical connections associated with GE MR Products. Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance… | Terminated |
| Aug 13, 2013 | Z-2291-2013GE Healthcare has recently become aware of a potential safety issue due to exposed yellow foam on the end bells of Discovery MR450, Discovery MR750, and Optima MR450w products. The yellow foam strip used between the end… | Terminated |
| Apr 9, 2013 | Z-2266-2013GE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products. For ClariView and Screen Save from GVTK (Gems Visualization… | Terminated |
| Jun 8, 2012 | Z-1207-2013GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO… | Terminated |
| Jul 29, 2011 | Z-3123-2011GE Healthcare has become aware of two potential safety issues associated with MR products (Optima MR450w, Discovery MR450, Discovery MR750, Brivo MR355, Optima MR360) that may impact patient safety. 1) Inline… | Terminated |
| May 10, 2011 | Z-0935-2013An error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports. | Terminated |
| Oct 6, 2009 | Z-0592-2010GE Healthcare has recently become aware that the use of Confirma, Inc.'s CADstream in conjunction with the GE Healthcare's PURE (Phased Array Uniformity Enhancement) software for dynamic MRI imaging (Signa Excite 1.5T,… | Terminated |
Other LNH Clearances
The newest 510(k)s for the same product code; their median review was 152 days; see every LNH device- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- MagicScan-1.5, GreenMR Serenity 1.5TK254250 · Xingaoyi Medical Equipment Co.,Ltd · Aug 17, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Oct 24, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files