Signa 3.0T MR750 System
System, Nuclear Magnetic Resonance Imaging. 510(k) K081028, from Ge Medical Systems, LLC (Waukesha, WI). Product code LNH, class 2.
As of Nov 19, 2026, K081028 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Apr 25, 2008 after 14 days, with 11 recalls tied to it in FDA's recall database.
- Review time: 14 days
- Recalls: 11
- Cited as a predicate by: 2 devices
The GE Signa® MR750 System is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The GE Signa® MR750 System is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and organs of the…
- GE 1.5T and 3.0T Signa Hdx Mr SystemK052293 · Ge Medical Systems, LLC · Sep 21, 2005
- Magnetom VerioK072237 · Siemens Medical Solutions USA, Inc. · Oct 10, 2007
- Discovery MR750W 3.0TK103327 · Ge Healthcare Japan Corporation · Sep 30, 2011
- Discovery MR450K083147 · Ge Medical Systems, LLC · Nov 4, 2008
| May 11, 2017 | Z-3025-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard. | Terminated |
| Dec 7, 2015 | Z-0201-2019GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup.… | Terminated |
| Apr 23, 2014 | Z-1533-2014GE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products. Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block… | Terminated |
| Apr 2, 2014 | Z-1480-2014Potential safety issue with gradient coil electromechanical connections associated with GE MR Products. Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance… | Terminated |
| Aug 13, 2013 | Z-2291-2013GE Healthcare has recently become aware of a potential safety issue due to exposed yellow foam on the end bells of Discovery MR450, Discovery MR750, and Optima MR450w products. The yellow foam strip used between the end… | Terminated |
| Apr 9, 2013 | Z-2266-2013GE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products. For ClariView and Screen Save from GVTK (Gems Visualization… | Terminated |
| Jun 8, 2012 | Z-1207-2013GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO… | Terminated |
| Jul 29, 2011 | Z-3122-2011GE Healthcare has become aware of two potential safety issues associated with MR products (Optima MR450w, Discovery MR450, Discovery MR750, Brivo MR355, Optima MR360) that may impact patient safety. 1) Inline… | Terminated |
| May 10, 2011 | Z-0935-2013An error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports. | Terminated |
| Nov 18, 2010 | Z-1375-2011GE Healthcare has become aware that during certain service procedures, there is a potential for electric shock associated with the gradient cooling manifold hoses of your MR750, MR450, and MR450w MR systems that may… | Terminated |
| Oct 6, 2009 | Z-0592-2010GE Healthcare has recently become aware that the use of Confirma, Inc.'s CADstream in conjunction with the GE Healthcare's PURE (Phased Array Uniformity Enhancement) software for dynamic MRI imaging (Signa Excite 1.5T,… | Terminated |
- MAGNETOM Free.XLK262011 · Siemens Healthineers AG · Nov 5, 2026
- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Apr 11, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files