Brivo MR355, Optima MR360

System, Nuclear Magnetic Resonance Imaging. 510(k) K103330, from Ge Hangwei Medical Systems Co., Ltd. (Beijing, CN). Product code LNH, class 2.

As of Nov 18, 2026, K103330 is an FDA 510(k) held by Ge Hangwei Medical Systems Co., Ltd., substantially equivalent on Feb 11, 2011 after 91 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 11, 2011
  • Review time: 91 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Brivo MR355/ Optima MR360 is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K103330, newest first; each opens FDA's recall record
Jun 14, 2024Z-2478-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE…Open, Classified
Jun 14, 2024Z-2477-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE…Open, Classified
Dec 23, 2021Z-0610-2022The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.Open, Classified
May 11, 2017Z-3021-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.Terminated
Jun 8, 2012Z-1209-2013GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO…Terminated
Nov 9, 2010Z-1433-2011GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the…Terminated

Other LNH Clearances

The newest 510(k)s for the same product code; their median review was 152 days; see every LNH device

Record

Type
Traditional
Received
Nov 12, 2010
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files