System, Nuclear Magnetic Resonance Imaging. 510(k) K121676, from Ge Medical Systems, LLC (Waukesha, WI). Product code LNH, class 2.

As of Nov 19, 2026, K121676 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Sep 20, 2012 after 105 days, with 10 recalls tied to it in FDA's recall database.

  • Review time: 105 days
  • Recalls: 10
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • EvryK200327 · Synaptive Medical, Inc. · Apr 29, 2020
  • Ziostation2K151212 · Qi Imaging, LLC · Nov 4, 2015

Recalls

Recalls FDA ties to K121676, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 14, 2024Z-2471-2024, Z-2463-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE…Open
Feb 18, 2022Z-0860-2022There is potential for the images to be flipped left to right.Open
Dec 23, 2021Z-0603-2022The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.Open
Aug 26, 2021Z-0141-2022Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.Terminated
Jun 4, 2021Z-2065-2021GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes…Open
Dec 23, 2019Z-0877-2020It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This…Terminated
Nov 15, 2018Z-0693-2019Possible incorrect software version loaded.Terminated
May 11, 2017Z-3028-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.Terminated
Apr 9, 2013Z-2266-2013GE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products. For ClariView and Screen Save from GVTK (Gems Visualization…Terminated

Other LNH clearances

The newest 510(k)s for the same product code; their median review was 143 days; see every LNH device

Record

Type
Traditional
Received
Jun 7, 2012
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files