GE Signa Excite 1.5T Mr System, GE Signa Excite 3.0T Mr System

System, Nuclear Magnetic Resonance Imaging. 510(k) K041476, from Ge Medical Systems, LLC (Milwaukee, WI). Product code LNH, class 2.

As of Nov 19, 2026, K041476 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Jun 17, 2004 after 14 days, with 15 recalls tied to it in FDA's recall database.

  • Review time: 14 days
  • Recalls: 15
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The GE Signa® Excite MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® Excite MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K041476, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 18, 2022Z-0858-2022There is potential for the images to be flipped left to right.Open
Dec 23, 2021Z-0615-2022The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.Open
Jun 7, 2018Z-0111-2019, Z-0112-2019, Z-0110-2019The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.Terminated
Mar 30, 2018Z-2527-2018There is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from…Terminated
Aug 31, 2015Z-0257-2016When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR…Terminated
Jul 10, 2015Z-2339-2015A gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with…Terminated
Mar 9, 2015Z-1305-2015GE Healthcare has become aware of a potential safety issue involving MRI systems due to software versions not being maintained properly at some sites.Terminated
Jan 7, 2015Z-1035-2015At certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The…Terminated
Nov 21, 2011Z-1943-2013Screen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the…Terminated
Nov 9, 2010Z-1429-2011GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the…Terminated
Oct 6, 2009Z-0592-2010GE Healthcare has recently become aware that the use of Confirma, Inc.'s CADstream in conjunction with the GE Healthcare's PURE (Phased Array Uniformity Enhancement) software for dynamic MRI imaging (Signa Excite 1.5T,…Terminated
Dec 14, 2007Z-1891-2008GE healthcare has become aware of an artifact which could affect the diagnostic capability associated with the 3.0T Propeller T2 Flair application of the 3.0T Signa HDx scanner.Terminated
Oct 1, 2007Z-2211-2008A software issue that may result in misregistration of functional and anatomical images associated with the 3.0T BrainwaveRT fMRI Application on the 3.0T Signa HDx scanner.Terminated

Other LNH clearances

The newest 510(k)s for the same product code; their median review was 143 days; see every LNH device

Record

Type
Special
Received
Jun 3, 2004
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files