GE Signa Excite 1.5T Mr System, GE Signa Excite 3.0T Mr System
System, Nuclear Magnetic Resonance Imaging. 510(k) K041476, from Ge Medical Systems, LLC (Milwaukee, WI). Product code LNH, class 2.
As of Nov 19, 2026, K041476 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Jun 17, 2004 after 14 days, with 15 recalls tied to it in FDA's recall database.
- Review time: 14 days
- Recalls: 15
- Cited as a predicate by: 4 devices
The GE Signa® Excite MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® Excite MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the…
- Signa Infinity Mr System With Excite TechnologyK013636 · GE Medical Systems · Jan 24, 2002
- GE 3.0T Signa Infinity Twinspeed With Excite Mr SystemK030874 · General Electric Co. · Mar 28, 2003
- Rthawk, Rthawk, Or Heartvista Workstation With RthawkK133848 · Heartvista, Inc. · Jun 25, 2014
- Vanguard Mri Auxiliary Patient Table With 4 and 8 Channel Coil Array for GE Signa SystemsK060873 · Sentinelle Medical, Inc. · Apr 14, 2006
- GE 1.5T and 3.0T Signa Hdx Mr SystemK052293 · Ge Medical Systems, LLC · Sep 21, 2005
- GE Signa Excite Mr Surgical OptionK043262 · GE Healthcare · Jan 12, 2005
| Feb 18, 2022 | Z-0858-2022There is potential for the images to be flipped left to right. | Open |
| Dec 23, 2021 | Z-0615-2022The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements. | Open |
| Jun 7, 2018 | Z-0111-2019, Z-0112-2019, Z-0110-2019The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch. | Terminated |
| Mar 30, 2018 | Z-2527-2018There is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from… | Terminated |
| Aug 31, 2015 | Z-0257-2016When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR… | Terminated |
| Jul 10, 2015 | Z-2339-2015A gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with… | Terminated |
| Mar 9, 2015 | Z-1305-2015GE Healthcare has become aware of a potential safety issue involving MRI systems due to software versions not being maintained properly at some sites. | Terminated |
| Jan 7, 2015 | Z-1035-2015At certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The… | Terminated |
| Nov 21, 2011 | Z-1943-2013Screen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the… | Terminated |
| Nov 9, 2010 | Z-1429-2011GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the… | Terminated |
| Oct 6, 2009 | Z-0592-2010GE Healthcare has recently become aware that the use of Confirma, Inc.'s CADstream in conjunction with the GE Healthcare's PURE (Phased Array Uniformity Enhancement) software for dynamic MRI imaging (Signa Excite 1.5T,… | Terminated |
| Dec 14, 2007 | Z-1891-2008GE healthcare has become aware of an artifact which could affect the diagnostic capability associated with the 3.0T Propeller T2 Flair application of the 3.0T Signa HDx scanner. | Terminated |
| Oct 1, 2007 | Z-2211-2008A software issue that may result in misregistration of functional and anatomical images associated with the 3.0T BrainwaveRT fMRI Application on the 3.0T Signa HDx scanner. | Terminated |
- MAGNETOM Free.XLK262011 · Siemens Healthineers AG · Nov 5, 2026
- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jun 3, 2004
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files