Innova 3100
Interventional Fluoroscopic X-Ray System. 510(k) K031637, from GE Medical Systems (Waukesha, WI). Product code OWB, class 2.
As of Nov 17, 2026, K031637 is an FDA 510(k) held by GE Medical Systems, substantially equivalent on Aug 22, 2003 after 87 days, with 25 recalls tied to it in FDA's recall database.
- Substantially equivalent: Aug 22, 2003
- Review time: 87 days
- Recalls: 25+
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Innova 3100 system is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, and optionally, rotational angiography procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Innova 4100K023178 · Ge Medical Systems, Inc. · Nov 26, 2002
- Innova And Innova 2000 SK022322 · Ge Medical Systems, Inc. · Oct 8, 2002
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ, Innova 2100IQ With Innova Ivus OptK061163 · Ge Medical Systems, LLC · Jun 7, 2006
- Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, And Innova 4100-IQK052412 · GE Healthcare · Sep 16, 2005
- Innova Ct Option For Digital Fluoroscopic Imaging Systems Innova Models 4100, 4100IQ, 3100, 3100IQ & 2100IQK052157 · Ge Medical Systems, LLC · Aug 25, 2005
- Innova 2100K050489 · Ge Medical Systems, LLC · Mar 11, 2005
- Innova 4100, Innova 3100K042053 · Ge Medical Systems, Inc. · Aug 26, 2004
Recalls
Recalls FDA ties to K031637, newest first; the 25 newest; each opens FDA's recall record| Dec 11, 2013 | Z-1163-2014GE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of… | Terminated |
| Jan 21, 2013 | Z-0947-2013A locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems. | Terminated |
| Jun 3, 2010 | Z-1922-2010GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. | Terminated |
| Jun 3, 2010 | Z-1923-2010GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. | Terminated |
| Jun 3, 2010 | Z-1921-2010GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. | Terminated |
| Dec 22, 2009 | Z-0618-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Jul 31, 2009 | Z-2241-2009GE Healthcare has recently became aware of a potential power supply failure with the Innova 2000, 3100 and 4100 systems that may impact patient safety. A potential power supply failure inside the GE equipment cabinet… | Terminated |
| Apr 16, 2009 | Z-0135-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Oct 9, 2008 | Z-1688-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Oct 9, 2008 | Z-1689-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Jul 25, 2008 | Z-2437-2008The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down. | Terminated |
| Jun 24, 2008 | Z-2356-2008GE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ… | Terminated |
| Jun 17, 2008 | Z-1475-2009GE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function.… | Terminated |
| May 19, 2008 | Z-1522-2008GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located… | Terminated |
| May 19, 2008 | Z-1521-2008GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located… | Terminated |
| Mar 6, 2008 | Z-1092-2009GE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset… | Terminated |
| Jan 30, 2008 | Z-0474-2009Software error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor… | Terminated |
| Jan 14, 2008 | Z-2040-2008Risk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system | Terminated |
| Mar 26, 2007 | Z-0869-2007Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital… | Terminated |
| Mar 26, 2007 | Z-0871-2007Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital… | Terminated |
| Mar 26, 2007 | Z-1044-2007Caution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage… | Terminated |
| Mar 26, 2007 | Z-0870-2007Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital… | Terminated |
| Mar 26, 2007 | Z-1043-2007Caution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage… | Terminated |
| Feb 5, 2007 | Z-0914-2007Unintended Gantry Movements : GE Healthcare has recently become aware of accidental, operated-initiated activation of gantry movements associated with the new, metallic Smart Box controller that may impact patient or… | Terminated |
| Mar 6, 2006 | Z-0897-06Incorrect screws were used for mounting the detector which may lead to a possible part falling onto a patient. | Terminated |
Other OWB Clearances
The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;EXTRON 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 27, 2003
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files