Innova 4100
Interventional Fluoroscopic X-Ray System. 510(k) K023178, from Ge Medical Systems, Inc. (Waukesha, WI). Product code OWB, class 2.
As of Nov 17, 2026, K023178 is an FDA 510(k) held by Ge Medical Systems, Inc., substantially equivalent on Nov 26, 2002 after 64 days, with 25 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 26, 2002
- Review time: 64 days
- Recalls: 25+
- Cited as a predicate by: 6 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Digital Fluoroscopic Imaging System is indicated for use in diagnostic and interventional angiographic procedures of human anatomy. It is intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures. This device is not intended for mammography applications.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Lcv+ Version 2K993037 · Ge Medical Systems, Inc. · Feb 14, 2000
- Multix Top/Pro Radiographic X-Ray TableK971452 · Siemens Medical Solutions USA, Inc. · May 14, 1997
Cited as a Predicate
6 later 510(k) summaries name this device as a predicate- Discovery Igs 740K133278 · GE Medical Systems SCS · Apr 24, 2014
- Canon, Model URS-50RFK093688 · Virtual Imaging, Inc. · Feb 4, 2010
- Canon Digital Fluoroscopy And Radiography, Model CXDI-50RFK092439 · Virtual Imaging, Inc. · Nov 30, 2009
- Centricity Radiology RA600, Cardiology CA1000 And Digital HardcopyK063628 · Ge Healthcare Integrated IT Solutions · Dec 26, 2006
- New BridK041780 · Medical Imaging Solutions, Inc. · Aug 11, 2004
- Innova 3100K031637 · GE Medical Systems · Aug 22, 2003
Recalls
Recalls FDA ties to K023178, newest first; the 25 newest; each opens FDA's recall record| Dec 11, 2013 | Z-1163-2014GE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of… | Terminated |
| Aug 16, 2013 | Z-0021-2014GE Healthcare has become aware of a potential safety issue involving missing screws in the overhead Video Monitor Suspension of the Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova… | Terminated |
| Jan 21, 2013 | Z-0947-2013A locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems. | Terminated |
| Jun 3, 2010 | Z-1922-2010GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. | Terminated |
| Jun 3, 2010 | Z-1923-2010GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. | Terminated |
| Jun 3, 2010 | Z-1921-2010GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. | Terminated |
| Dec 22, 2009 | Z-0619-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Aug 6, 2009 | Z-0392-2011GE Healthcare has become aware of a potential video splitter failure with certain Innova 2000 and 4100 X-ray systems that may impact patient safety. A potential video splitter failure inside the GE equipment cabinet… | Terminated |
| Apr 16, 2009 | Z-0137-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Oct 9, 2008 | Z-1690-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Oct 9, 2008 | Z-1691-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Jul 25, 2008 | Z-2438-2008The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down. | Terminated |
| Jun 24, 2008 | Z-2358-2008GE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ… | Terminated |
| Jun 17, 2008 | Z-1476-2009GE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function.… | Terminated |
| May 19, 2008 | Z-1522-2008GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located… | Terminated |
| Mar 6, 2008 | Z-1093-2009GE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset… | Terminated |
| Jan 30, 2008 | Z-0475-2009Software error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor… | Terminated |
| Jan 14, 2008 | Z-2041-2008Risk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system | Terminated |
| Mar 26, 2007 | Z-0869-2007Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital… | Terminated |
| Mar 26, 2007 | Z-0871-2007Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital… | Terminated |
| Mar 26, 2007 | Z-1044-2007Caution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage… | Terminated |
| Mar 26, 2007 | Z-0870-2007Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital… | Terminated |
| Mar 26, 2007 | Z-1043-2007Caution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage… | Terminated |
| Feb 5, 2007 | Z-0915-2007Unintended Gantry Movements : GE Healthcare has recently become aware of accidental, operated-initiated activation of gantry movements associated with the new, metallic Smart Box controller that may impact patient or… | Terminated |
| Mar 6, 2006 | Z-0898-06Incorrect screws were used for mounting the detector which may lead to a possible part falling onto a patient. | Terminated |
Other OWB Clearances
The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;EXTRON 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 23, 2002
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files