Lcv+ Version 2

Interventional Fluoroscopic X-Ray System. 510(k) K993037, from Ge Medical Systems, Inc. (Waukesha, WI). Product code OWB, class 2.

As of Nov 18, 2026, K993037 is an FDA 510(k) held by Ge Medical Systems, Inc., substantially equivalent on Feb 14, 2000 after 158 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 14, 2000
  • Review time: 158 days
  • Recalls: 6
  • Cited as a predicate by: 5 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The LCV+ Version 2 is indicated for use in generating fluoroscopic images of human anatomy for cardiology diagnostic/interventional procedures. It is intended to replace fluoroscopic images obtained through the image intensifier technology.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K993037, newest first; each opens FDA's recall record
Aug 16, 2013Z-0021-2014GE Healthcare has become aware of a potential safety issue involving missing screws in the overhead Video Monitor Suspension of the Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova…Terminated
Jun 10, 2010Z-2238-2010A hardware component (potentiometer and encoder)mismatch in Innova 2000 digital fluoroscopic imaging systems may cause a Source to Image Distance (SID) failure, which will cause the imager to fail to detect the distance…Terminated
Dec 22, 2009Z-0620-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some…Terminated
Aug 6, 2009Z-0392-2011GE Healthcare has become aware of a potential video splitter failure with certain Innova 2000 and 4100 X-ray systems that may impact patient safety. A potential video splitter failure inside the GE equipment cabinet…Terminated
Aug 6, 2009Z-0393-2011GE Healthcare has become aware of a potential video splitter failure with certain Innova 2000 and 4100 X-ray systems that may impact patient safety. A potential video splitter failure inside the GE equipment cabinet…Terminated
Jul 25, 2008Z-2435-2008The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.Terminated

Other OWB Clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Sep 9, 1999
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files