Multix Top/Pro Radiographic X-Ray Table
Table, Radiographic, Non-Tilting, Powered. 510(k) K971452, from Siemens Medical Solutions USA, Inc. (Iselin, NJ). Product code IZZ, class 2.
As of Nov 18, 2026, K971452 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on May 14, 1997 after 23 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 23 days
- Recalls: 8
- Cited as a predicate by: 3 devices
- Multix U. Bucky Table X-Ray UnitK803130 · Siemens Corp. · Jan 28, 1981
- Compax Model 40, 400, 400TK884930 · General Electric Co. · Dec 22, 1988
- Dx-SiK063421 · Agfa Corp. · Nov 22, 2006
- Innova 4100K023178 · Ge Medical Systems, Inc. · Nov 26, 2002
- Fuji Speed Suite, Fuji Speed Sute Phase IiK013218 · Fujifilm Medical System U.S.A., Inc. · Dec 20, 2001
| May 31, 2024 | Z-2183-2024, Z-2181-2024, Z-2186-2024, Z-2182-2024, Z-2187-2024, Z-2180-2024Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire | Open |
| Aug 6, 2018 | Z-0298-2019In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to… | Terminated |
| Mar 5, 2014 | Z-1460-2014There is a potential issue and possible hazard to patients when using the AXIOM Vertix MD Trauma systems. In rare cases, steel ropes inside the lift column of the system can be defective without triggering the safety… | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 21, 1997
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1980
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files