Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, And Innova 4100-IQ
Interventional Fluoroscopic X-Ray System. 510(k) K052412, from GE Healthcare (Waukesha, WI). Product code OWB, class 2.
As of Nov 17, 2026, K052412 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Sep 16, 2005 after 14 days, with 25 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 16, 2005
- Review time: 14 days
- Recalls: 25+
- Cited as a predicate by: 7 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFFor Innova 4100, Innova 4100$^{IQ}$, Innova 3100, Innova 3100$^{IQ}$, Innova 2100$^{IQ}$ with Innova 3D or InnovaSpin option: The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Advantage 3D XrK974715 · GE Medical Systems · Jul 6, 1998
- Innova 2100K050489 · Ge Medical Systems, LLC · Mar 11, 2005
- Digital Fluoroscopic Imaging System, Model Innova 4100 With Tilt Table OptionK033244 · Ge Medical Systems, Inc. · Oct 16, 2003
- Innova 3100K031637 · GE Medical Systems · Aug 22, 2003
Cited as a Predicate
7 later 510(k) summaries name this device as a predicate- Stereo 3D option for Vision ApplicationsK152352 · GE Medical Systems SCS · Jan 20, 2016
- Ge Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model: 4100IQ, 3100IQ, 2100IQK092004 · GE Healthcare · Aug 25, 2009
- Ge Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Models Innova 4100, 3100, 2100, 3131, 2121K091658 · GE Healthcare · Jun 29, 2009
- INFX-8000F (Infinix Csi And Infinix Dpi)K081624 · Toshiba America Medical Systems, In.C · Jun 23, 2008
- INFX-8000C (Infinix Cc-I And Vc-I)K081621 · Toshiba America Medical Systems, In.C · Jun 23, 2008
- Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ, Innova 2100IQ With Innova Ivus OptK061163 · Ge Medical Systems, LLC · Jun 7, 2006
- Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121K060259 · Ge Medical Systems, LLC · Feb 22, 2006
Recalls
Recalls FDA ties to K052412, newest first; the 25 newest; each opens FDA's recall record| Sep 4, 2014 | Z-0453-2015Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism. | Terminated |
| Dec 20, 2011 | Z-1318-2013While performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB).… | Terminated |
| Oct 21, 2011 | Z-1142-2013As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In… | Terminated |
| Aug 24, 2011 | Z-0133-2012A potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails. | Terminated |
| May 16, 2011 | Z-1616-2013GE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems. | Terminated |
| Feb 4, 2011 | Z-0934-2013Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected. | Terminated |
| Mar 24, 2010 | Z-1659-2010A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there… | Terminated |
| Mar 24, 2010 | Z-1661-2010A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there… | Terminated |
| Mar 24, 2010 | Z-1660-2010A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there… | Terminated |
| Dec 22, 2009 | Z-0619-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Dec 22, 2009 | Z-0617-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Dec 22, 2009 | Z-0618-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Aug 5, 2009 | Z-0408-2011GE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy… | Terminated |
| Aug 5, 2009 | Z-0409-2011GE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy… | Terminated |
| Aug 5, 2009 | Z-0410-2011GE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy… | Terminated |
| Apr 16, 2009 | Z-0135-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Apr 16, 2009 | Z-0137-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Apr 16, 2009 | Z-0136-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Nov 4, 2008 | Z-1036-2009One of three copper spectral filters could remain within the x-ray while the system displays no spectral filter. The displayed Air Kerma rate and thus cumulative Air Kerma are overestimated. The Air Kerma rate may be… | Terminated |
| Oct 9, 2008 | Z-1690-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Oct 9, 2008 | Z-1691-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Oct 9, 2008 | Z-1688-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Oct 9, 2008 | Z-1689-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Oct 9, 2008 | Z-1687-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
| Jul 25, 2008 | Z-2436-2008The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down. | Terminated |
Other OWB Clearances
The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;EXTRON 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Sep 2, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files