GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model: 4100IQ, 3100IQ, 2100IQ
Interventional Fluoroscopic X-Ray System. 510(k) K092004, from GE Healthcare (Waukesha, WI). Product code OWB, class 2.
As of Nov 18, 2026, K092004 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Aug 25, 2009 after 54 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 54 days
- Recalls: 2
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQK052412 · GE Healthcare · Sep 16, 2005
- Innova 2100K050489 · Ge Medical Systems, LLC · Mar 11, 2005
- Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121K060259 · Ge Medical Systems, LLC · Feb 22, 2006
- Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ, Innova 2100IQ With Innova Ivus OptK061163 · Ge Medical Systems, LLC · Jun 7, 2006
- The Ic-Pro System, Model Version 3.2K083745 · Paieon, Inc. · Feb 17, 2009
- Philips Fresco Release 1K031836 · Philips Medical Systems North America Co. · Aug 14, 2003
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Innova Vision, Model S18751VN, Innova Trackvision, Model S18751VT, Innova Epvision, Model S18751EPK092639 · GE Medical Systems SCS · Dec 2, 2009
Recalls
Recalls FDA ties to K092004, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Aug 24, 2011 | Z-0133-2012A potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails. | Terminated |
| Feb 4, 2011 | Z-0934-2013Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected. | Terminated |
Other OWB clearances
The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;Extron 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 2, 2009
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files