Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ, Innova 2100IQ With Innova Ivus Opt
Interventional Fluoroscopic X-Ray System. 510(k) K061163, from Ge Medical Systems, LLC (Waukesha, WI). Product code OWB, class 2.
As of Nov 19, 2026, K061163 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Jun 7, 2006 after 42 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 42 days
- Recalls: 25+
- Cited as a predicate by: 3 devices
For Innova 2121 $^{IQ}$, Innova 3131 $^{IQ}$, Innova 4100, Innova 4100$^{IQ}$, Innova 3100, Innova 3100$^{IQ}$, Innova 2100$^{IQ}$ devices with Innova IVUS option: The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures.
- Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121K060259 · Ge Medical Systems, LLC · Feb 22, 2006
- Digital Fluoroscopic Imaging System, Model Innova 4100 With Tilt Table OptionK033244 · Ge Medical Systems, Inc. · Oct 16, 2003
- Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQK052412 · GE Healthcare · Sep 16, 2005
- Innova 3100K031637 · GE Medical Systems · Aug 22, 2003
- Innova 2100K050489 · Ge Medical Systems, LLC · Mar 11, 2005
- Integrated Innova - S5I System OptionK111209 · GE Healthcare · Jul 26, 2011
- GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model: 4100IQ, 3100IQ, 2100IQK092004 · GE Healthcare · Aug 25, 2009
- GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Models Innova 4100, 3100, 2100, 3131, 2121K091658 · GE Healthcare · Jun 29, 2009
| Aug 16, 2012 | Z-2376-2012GE Healthcare has recently become aware of a potential safety issue with respect to the Innova 2121IQ & Innova 3131IQ Systems. While performing a fluoroscopic acquisition with the Innova 2121IQ or Innova 3131IQ systems,… | Terminated |
| Dec 20, 2011 | Z-1318-2013While performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB).… | Terminated |
| Aug 24, 2011 | Z-0133-2012A potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails. | Terminated |
| May 16, 2011 | Z-1616-2013GE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems. | Terminated |
| Feb 4, 2011 | Z-0934-2013Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected. | Terminated |
| Mar 24, 2010 | Z-1659-2010, Z-1661-2010, Z-1660-2010A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there… | Terminated |
| Dec 22, 2009 | Z-0621-2010, Z-0619-2010, Z-0617-2010, Z-0618-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Aug 5, 2009 | Z-0408-2011, Z-0409-2011, Z-0407-2011, Z-0410-2011GE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy… | Terminated |
| Apr 16, 2009 | Z-0138-2010, Z-0135-2010, Z-0137-2010, Z-0136-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Nov 21, 2008 | Z-1195-2009GE Healthcare has become aware of a potential image storage failure on Innova 2121 IQ/ 3131 IQ cardiovascular imaging systems that may impact patient safety. This intermittent communication failure on the image storage… | Terminated |
| Oct 9, 2008 | Z-1690-2009, Z-1691-2009, Z-1688-2009, Z-1689-2009GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration… | Terminated |
- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;Extron 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 26, 2006
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files