INFX-8000F (Infinix Csi and Infinix Dpi)

Solid State X-Ray Imager (Flat Panel/Digital Imager). 510(k) K081624, from Toshiba America Medical Systems, In.C (Tustin, CA). Product code MQB, class 2.

As of Nov 19, 2026, K081624 is an FDA 510(k) held by Toshiba America Medical Systems, In.C, substantially equivalent on Jun 23, 2008 after 13 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 13 days
  • Recalls: 8
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

This device is a digital radiography/fluoroscopy system used in a diagnostic angiography configuration. This system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K081624, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 5, 2021Z-1176-2021The x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.Open
Dec 8, 2020Z-1271-2021System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.Open
Mar 3, 2020Z-2327-2020Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.Open
Feb 8, 2019Z-0360-2020Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.Terminated
Dec 29, 2017Z-0842-2018During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.Terminated
Nov 8, 2012Z-0994-2013Lost and incorrect images. When using the "F-Rec" fluoroscopy image acquisition recording option in the "manual mode" an infrequent software error may cause images from another study to be displayed on the monitor. The…Terminated
Oct 3, 2011Z-0331-2012The recall was initiated because Toshiba has confirmed that during patient transfer from a gurney to the catherization table, the tabletop may rotate unexpectedly when sufficient force is applied to the tabletop. The…Terminated
Sep 29, 2008Z-0005-2009Audible signal is not available at all times when using the HLC fluoroscopy mode. Special means of activation are not designed in for selection of High Level Controls, as required by regulation.Terminated

Other MQB clearances

The newest 510(k)s for the same product code; their median review was 179 days; see every MQB device

Record

Type
Traditional, third-party review
Received
Jun 10, 2008
Review panel
Radiology
Regulation
21 CFR 892.1680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files