Alphenix, INFX-8000F/B, V8.0

Interventional Fluoroscopic X-Ray System. 510(k) K182415, from Canon Medical Systems Corporation (Otawara-Shi, JP). Product code OWB, class 2.

As of Nov 19, 2026, K182415 is an FDA 510(k) held by Canon Medical Systems Corporation, substantially equivalent on Sep 28, 2018 after 24 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 24 days
  • Recalls: 5

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K182415, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 12, 2026Z-0290-2027Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.Open
Aug 4, 2026Z-3014-2026Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.Open
Dec 19, 2025Z-1177-2026It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result,…Open
Nov 15, 2023Z-1155-2024VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead…Open
Dec 20, 2021Z-0492-2022Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a…Open

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Sep 4, 2018
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files