Arca
Skip to content
Tools
Tools

Alphenix, INFX-8000V/B, V8.0

Interventional Fluoroscopic X-Ray System. 510(k) K181670, from Canon Medical Systems Corporation (Otawara-Shi, JP). Product code OWB, class 2.

As of Nov 19, 2026, K181670 is an FDA 510(k) held by Canon Medical Systems Corporation, substantially equivalent on Jul 25, 2018 after 30 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

This device is a digital radiography/fluoroscopy system used in a diagnostic interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K181670, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Dec 20, 2021Z-0492-2022Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a…Open

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Jun 25, 2018
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.