Innova Ct Option For Digital Fluoroscopic Imaging Systems Innova Models 4100, 4100IQ, 3100, 3100IQ & 2100IQ
System, X-Ray, Tomography, Computed. 510(k) K052157, from Ge Medical Systems, LLC (Waukesha, WI). Product code JAK, class 2.
As of Nov 17, 2026, K052157 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Aug 25, 2005 after 17 days, with 25 recalls tied to it in FDA's recall database.
- Substantially equivalent: Aug 25, 2005
- Review time: 17 days
- Recalls: 25+
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFFor Innova 4100, Innova 4100$^{IQ}$, Innova 3100, Innova 3100$^{IQ}$, Innova 2100$^{IQ}$ with Innova 3T option: The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- DynactK042646 · Siemens Medical Systems, Inc. · Oct 19, 2004
- Digital Fluoroscopic Imaging System, Model Innova 4100 With Tilt Table OptionK033244 · Ge Medical Systems, Inc. · Oct 16, 2003
- Innova 3100K031637 · GE Medical Systems · Aug 22, 2003
Recalls
Recalls FDA ties to K052157, newest first; the 25 newest; each opens FDA's recall record| Jan 21, 2013 | Z-0947-2013A locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems. | Terminated |
| Dec 20, 2011 | Z-1318-2013While performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB).… | Terminated |
| Oct 21, 2011 | Z-1142-2013As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In… | Terminated |
| Aug 24, 2011 | Z-0133-2012A potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails. | Terminated |
| May 16, 2011 | Z-1616-2013GE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems. | Terminated |
| Mar 24, 2010 | Z-1659-2010A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there… | Terminated |
| Mar 24, 2010 | Z-1661-2010A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there… | Terminated |
| Mar 24, 2010 | Z-1660-2010A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there… | Terminated |
| Dec 22, 2009 | Z-0619-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Dec 22, 2009 | Z-0617-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Dec 22, 2009 | Z-0618-2010Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some… | Terminated |
| Aug 5, 2009 | Z-0408-2011GE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy… | Terminated |
| Aug 5, 2009 | Z-0409-2011GE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy… | Terminated |
| Aug 5, 2009 | Z-0410-2011GE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy… | Terminated |
| Apr 16, 2009 | Z-0135-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Apr 16, 2009 | Z-0137-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Apr 16, 2009 | Z-0136-2010Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message… | Terminated |
| Jul 25, 2008 | Z-2436-2008The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down. | Terminated |
| Jul 25, 2008 | Z-2438-2008The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down. | Terminated |
| Jul 25, 2008 | Z-2437-2008The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down. | Terminated |
| Jun 17, 2008 | Z-1476-2009GE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function.… | Terminated |
| Jun 17, 2008 | Z-1475-2009GE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function.… | Terminated |
| Jun 17, 2008 | Z-1474-2009GE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function.… | Terminated |
| Jan 30, 2008 | Z-0474-2009Software error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor… | Terminated |
| Jan 30, 2008 | Z-0473-2009Software error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor… | Terminated |
Other JAK Clearances
The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 8, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files