Arca
Skip to content
Tools
Tools

PE Centrifuge Temperature Controlled

Recall Z-2980-2024, part of FDA recall event 95064. See Beckman Coulter Biomedical's company page.

Beckman Coulter Biomedical GmbH started this Class II device recall on Jul 17, 2024 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Jul 17, 2024
  • Status: Ongoing
  • Products in this event: 3

Why It Was Recalled

Class II

The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.

Record

Product
PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
Quantity
—
Lots and codes
UDI: (01)15099590673567/ All serial numbers
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.
Recalling firm
Beckman Coulter Biomedical GmbH, Munchen, Germany
How the firm notified
Letter
FDA classified
Sep 5, 2024

Other Products in This Recall

Recall event 95064

Beckman Coulter Biomedical GmbH's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports