DxA 5000; Catalog Number:B50516; UDI:15099590707118 - Product Usage: The DxA 5000 is a…
Recall Z-2983-2020, part of FDA recall event 85874. See Beckman Coulter Biomedical's company page.
Beckman Coulter Biomedical GmbH started this Class II device recall on Jun 17, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 24, 2022.
- Classification: Class II
- Started: Jun 17, 2020
- Status: Terminated
The DxA 5000 instructions for use may be confusing regarding the re-use of sample IDs. Beckman Coulter has received complaints where the re-use of sample barcodes were not being detected by the system .
- Product
- DxA 5000; Catalog Number:B50516; UDI:15099590707118 - Product Usage: The DxA 5000 is a high-speed, modular, automated sample handling system that performs pre-analytical and post-analytical sample processing and storage. The automation system also sorts, routes, and presents sample tubes to analyzers for analysis. The DxA 5000 also consolidates a variety of analytical instruments, such as an Immunoassay analyzer, into a unified workstation on a track system.
- Quantity
- 32 units
- Lots and codes
- Serial/Lot Number(s)/Software Version: all
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of Illinois, Ohio, Pennsylvania, and West Virginia. The countries of Australia, Canada, China, Czech Republic, France, Israel, Italy, Germany, Switzerland, Spain, Netherlands, and United Kingdom.
- Recalling firm
- Beckman Coulter Biomedical GmbH, Munich, N/A, Germany
- How the firm notified
- Letter
- FDA classified
- Sep 18, 2020
Newest first, one product per recall event; search them all
- REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager…Z-2834-2026 · Class II · Apr 29, 2026 · Ongoing
- PE Centrifuge, REF: B36365, a component of the Power ExpressZ-2979-2024 · Class II · Jul 17, 2024 · Ongoing