Beckman Coulter DxA Automation System
Recall Z-2144-2023, part of FDA recall event 92484. See Beckman Coulter Biomedical's company page.
Beckman Coulter Biomedical GmbH started this Class II device recall on Jun 2, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jun 2, 2023
- Status: Ongoing
- Products in this event: 2
A software defect in the DxA Automation System may cause tests to be run on patient samples in tube types not compatible with the requested test, leading to erroneous patient results
- Product
- Beckman Coulter DxA Automation System, DxA 5000, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnly
- Quantity
- 146 systems
- Lots and codes
- UDI-DI: (01)15099590707118, All DxA serial numbers
- Distribution
- US: AL CA FL IL IN KY MI MN MO NE NJ NV NY OH PA SC TX WI WV OUS: Australia, Belgium, Canada, China, Croatia, Czechia, Egypt, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Saudi Arabia, Spain, Switzerland, United Kingdom,
- Recalling firm
- Beckman Coulter Biomedical GmbH, Munchen, N/A, Germany
- How the firm notified
- Letter
- FDA classified
- Jul 14, 2023
- Beckman Coulter DxA Automation System, DxA 5000 Fit, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnlyZ-2145-2023 · Class II · Jun 2, 2023 · Ongoing
Newest first, one product per recall event; search them all
- REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager…Z-2834-2026 · Class II · Apr 29, 2026 · Ongoing
- PE Centrifuge, REF: B36365, a component of the Power ExpressZ-2979-2024 · Class II · Jul 17, 2024 · Ongoing
- DxA 5000; Catalog Number:B50516; UDI:15099590707118 - Product Usage: The DxA 5000 is a high-speed, modular, automated…Z-2983-2020 · Class II · Jun 17, 2020 · Terminated