Arca
Skip to content
Tools
Tools

REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info…

Recall Z-2834-2026, part of FDA recall event 98927. See Beckman Coulter Biomedical's company page.

Beckman Coulter Biomedical GmbH started this Class II device recall on Apr 29, 2026 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Apr 29, 2026
  • Status: Ongoing

Why It Was Recalled

Class II

Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.

Record

Product
REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections
Quantity
20 systems
Lots and codes
Software Version 2.5 UDI-DI code: 15099590802073
Distribution
U.S.: FL, GA, IL, MI, and PA O.U.S.: Belgium, France, Hungary, Poland, Taiwan (Province of China), Tunisia, United Arab Emirates, and United Kingdom of Great Britain and Northern Ireland.
Recalling firm
Beckman Coulter Biomedical GmbH, Munchen, Germany
How the firm notified
Letter
FDA classified
Jul 31, 2026

Beckman Coulter Biomedical GmbH's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports